For Research Use Only. Not for use in diagnostic procedures.
QuantStudio
3 and 5 RealTime PCR
Systems
INSTALLATION, USE, AND MAINTENANCE
Firmware v1.6
Publication Number MAN0010407
Revision F
Life Technologies Holdings Pte Ltd | Block 33 | Marsiling Industrial Estate Road 3 | #07-06, Singapore 739256
For descriptions of symbols on product labels or product documents, go to thermofisher.com/symbols-definition.
Revision history: MAN0010407 F (English)
Revision Date Description
F 8 August 2024
Information about using tubes/tube strips was added (“Options for loading MicroAmp
consumables with
the QuantStudio
3 and 5 RealTime PCR Systems” on page40 and “Prepare samples using MicroAmp
tubes/tube strips with separate cap strips” on page43).
Caution information about flat caps for tubes was moved to “Prepare samples using MicroAmp
tubes/tube
strips with separate cap strips” on page43.
Information about instrument power issues during runs was added (“View run history and delete or transfer
files from the instrument” on page47).
Minor verbiage was updated.
E.0 31 January 2024
v1.6
The information about instrument settings was updated to indicate that this feature is available only to users
that are signed in when SAE mode is enabled (“Configure instrument settings” on page27).
The instructions to delete a run from the run history were updated to indicate that only an Administrator
can perform this function when SAE mode is enabled (
“View run history and delete or transfer files from the
instrument” on page47).
A data management feature was added to the settings (“Configure instrument settings” on page27).
Instructions were added to perform a self-verification test, to view the instrument logs, and to connect the
instrument to a network drive.
The instructions to set the date and time were updated. The date and time cannot be set if the instrument is
linked to the Thermo Fisher
Connect Platform.
The instructions to test the connection to the Thermo Fisher
Connect Platform were removed. This feature
is not available.
The requirements for the computer for the desktop software were updated.
The support that is provided for the computer for the desktop software was updated.
D.0 21 July 2021
Updates – describes new features in Firmware v1.4.x and other changes, including:
Updated the name of the cloud-based platform to Thermo Fisher
Connect.
Added instructions for connecting to the Thermo Fisher
Connect cloud-based platform.
Updated the firewall instrument-to-computer port number.
Updated licensing information.
Added Network and password security requirements section
Added Perform internal validations before software upgrades section
Added information for updating the firmware.
Update operating system to Windows
10 (64bit).
Added the MicroAmp
Optical Film Compression Pad as an optional accessory for the 96-well 0.2-mL block
when using a MicroAmp
Optical 96-Well Reaction Plate, including instructions to load and unload the plate
if using this accessory.
Specified that an instrument profile is only for the instrument. It is not related to any other account for the
system or the software.
Specified that if the connection between the instrument and the desktop software is interrupted during a run,
the instrument will complete the run. The EDS
file must be transferred manually to the desktop software with
a USB drive or through a network drive.
Changed all references to Bleach, 10% solution to Na-hypochlorite (0.1% v/v) solution
Revision Date Description
D.0
(contin
ued)
21 July 2021
Updated the calibration and verification schedule.
Specified that Background and Dye calibrations must be performed after an ROI/Uniformity
calibration.
Specified ROI/Uniformity calibration invalidates all other calibrations.
Added that the Background calibration can be used to obtain the most accurate data for the removal
of background fluorescence.
Specified that a Background calibration and a Dye calibration do not invalidate all other calibrations.
Added a recommendation that RNase P instrument verification be performed after instrument
calibrations.
Added troubleshooting information for RNase P verification.
Added troubleshooting information if the connection between the instrument and the computer is not
recognized.
Added storage and transport conditions to the environmental requirements.
Specified that restoring factory defaults can only be performed by an administrator and specified that the
most recent valid ROI/Uniformity, Background, and System Dye calibrations are not deleted when factory
defaults are restored.
C.0 16 December
2015
Updates – describes new features in Firmware v1.2.x, including:
Updated workflow to create an instrument profile
Streamlined workflow for connectivity to Thermo Fisher
Connect
Improved display of VeriFlex
Zones
Plate insert reminder before starting run
Support for 96-well Fast (0.1 mL) plates
B.0 1 October 2015
Updates – describes new features in Firmware v1.1.x, including:
Experiment runs: monitor real-time data, edit cycle number and lock screen during a run, view end plot, and
support for 384-well plates
File management: access Network folders, navigate folder structures and save templates on instrument,
and perform batch actions for
file management
Instrument configuration: support for Security, Audit, and e-Signature (SAE), smart monitoring, and choice
of server region
Includes information on software feature comparison, definitions of terms, parts of method, network connection
options, experiment types, desktop software installation.
A.0 23 April 2015
New document. Describes installation, operation, and maintenance of the QuantStudio
3 and 5 RealTime PCR
Systems.
The information in this guide is subject to change without notice.
DISCLAIMER: TO THE EXTENT ALLOWED BY LAW, THERMO FISHER SCIENTIFIC INC. AND/OR ITS AFFILIATE(S) WILL NOT BE
LIABLE FOR SPECIAL, INCIDENTAL, INDIRECT, PUNITIVE, MULTIPLE, OR CONSEQUENTIAL DAMAGES IN CONNECTION WITH OR
ARISING FROM THIS DOCUMENT, INCLUDING YOUR USE OF IT.
Important Licensing Information: These products may be covered by one or more Limited Use Label Licenses. By use of these
products, you accept the terms and conditions of all applicable Limited Use Label Licenses.
Trademarks: All trademarks are the property of Thermo Fisher Scientific and its subsidiaries unless otherwise specified. TaqMan is
a trademark of Roche Molecular Systems, Inc., used under permission and license. Cy is a registered trademark of GE Healthcare.
Windows is a trademark of Microsoft Corporation. Chrome is a trademark of Google LLC. Core and Intel are trademarks of Intel
Corporation.
©2015-2024 Thermo Fisher Scientific Inc. All rights reserved.
Contents
About thisguide .......................................................... 10
Purpose ...................................................................... 10
CHAPTER1Productinformation ................................................ 11
Network and password security requirements ..................................... 11
Network configuration andsecurity .......................................... 11
Passwordsecurity ......................................................... 11
Perform internal validations before software upgrades .............................. 11
Update the instrumentfirmware .................................................. 12
Back up theinstrument ..................................................... 12
Instrument hardwaredescription ................................................. 12
Instrument overview ....................................................... 12
Parts of theinstrument ..................................................... 16
Softwaredescription ........................................................... 16
Instrument, desktop, and cloud software features .............................. 16
Folders, templates, experiments, and projects ................................. 17
Third-party software ....................................................... 19
Network connectionoptions ..................................................... 19
Types ofruns .................................................................. 19
CHAPTER2Start, sign on, and configure theinstrument ..................... 22
Installation and instrumentverification ............................................ 22
Precautions foruse ............................................................. 23
Power on theinstrument ........................................................ 23
Parts of the home screen ....................................................... 24
Use the instrument without signingin ............................................. 26
Create a new instrumentprofile .................................................. 26
Signin ........................................................................ 27
Signout ...................................................................... 27
Configure instrumentsettings .................................................... 27
4
QuantStudio
3 and 5 RealTime PCR Systems Installation, Use, and Maintenance Guide
CHAPTER3Create and run experiments on theinstrument .................. 31
Workflow ..................................................................... 32
Options for running an experiment ............................................... 33
Create and run an experiment from a template ................................ 33
Run an experiment from a savedfile ......................................... 33
Repeat the last instrumentrun ............................................... 34
Edit an experiment before starting arun .......................................... 35
Enter or edit template properties ............................................. 35
Edit the runmethod ........................................................ 36
Define, assign, and view welldetails ......................................... 39
Options for loading MicroAmp
consumables with the QuantStudio
3 and 5
RealTime PCR Systems ...................................................... 40
Load and unload the plate ...................................................... 40
Load and unload a plate in theinstrument .................................... 40
(If applicable) load and unload a plate in the instrument with a compressionpad ... 41
Prepare samples using MicroAmp
tubes/tube strips with separate capstrips ......... 43
View, pause, or stop arun ....................................................... 45
View real-time data and plots on the instrument touchscreen .................... 45
Pause or stop an instrumentrun ............................................. 45
Adjust the display of real-time plots on the instrument touchscreen .............. 46
Lock the touchscreen during arun ........................................... 46
Transfer, view, or manage files and results ......................................... 46
Transfer EDS files from the instrument home screen ............................ 46
View run history and delete or transfer files from theinstrument ................. 47
Manage templates (EDTfiles) ................................................ 47
CHAPTER4Calibrate and verify instrument performance ..................... 49
Calibration and verificationschedule ............................................. 49
Calibrationdescriptions ......................................................... 50
View the calibration status and set reminders ...................................... 51
View calibration status and set reminders in theinstrument ..................... 51
View calibration status and set reminders in Thermo Fisher
ConnectPlatform .... 51
Perform ROI/uniformity, background, and dye calibrations .......................... 52
Workflow: Calibration ...................................................... 52
Prepare a calibration plate .................................................. 53
Perform calibrations ........................................................ 54
View calibration images and transfer results to a USBdrive ..................... 56
Troubleshoot calibration failure .............................................. 57
Identifycontamination ...................................................... 57
Create a background plate (optional) ......................................... 58
Contents
QuantStudio
3 and 5 RealTime PCR Systems Installation, Use, and Maintenance Guide
5
Perform instrument verification using RNase P plates ............................... 59
Instrument verificationdescription ........................................... 59
RNaseP instrument verification plate ........................................ 59
Performance specifications pass criteria ...................................... 61
Prepare an RNase P plate ................................................... 61
Perform RNasePverification ................................................ 62
Troubleshoot verification failure .............................................. 63
Calibrate customdyes .......................................................... 65
Custom dyes overview ..................................................... 65
Use a dilution series to determine an optimal custom dye concentration .......... 65
Calibrate the customdye ................................................... 68
Calibrate for a custom melt curve experiment ...................................... 72
CHAPTER5Maintain theinstrument ............................................ 74
Back up or restore theinstrument ................................................ 74
Perform a self-verification test ................................................... 75
View instrumentlogs ........................................................... 75
Configure the networkdrive ..................................................... 75
Create a shared networkfolder .............................................. 76
Determine the IP address of the computer .................................... 77
Determine the DNSname ................................................... 77
Decontaminate the sampleblock ................................................ 77
Materials required .......................................................... 77
Clean the sampleblock ..................................................... 78
Detailed procedures for cleaning the sampleblock ............................. 79
Replace the instrumentfuses .................................................... 80
Materials required .......................................................... 80
Replace thefuses .......................................................... 80
Power on or o, store, andmove ................................................. 81
Enable sleepmode ........................................................ 81
Power on theinstrument .................................................... 81
Power o theinstrument .................................................... 82
Prepare the instrument to store, move, orship ................................. 82
Move theinstrument ....................................................... 82
Return the instrument for service ............................................ 83
Contents
6
QuantStudio
3 and 5 RealTime PCR Systems Installation, Use, and Maintenance Guide
CHAPTER6Configure the instrument and manage instrumentprofiles ..... 84
Initial startup .................................................................. 84
Manageprofiles ................................................................ 85
Create an administrator instrument profile during initial start-up ................. 85
Manage all instrument profiles (Administratoronly) ............................. 85
Edit an instrumentprofile ................................................... 86
Link an instrument profile to Thermo Fisher
Connect .............................. 86
Recommended order to set upprofiles ....................................... 87
Overview of local instrument profiles and Thermo Fisher
Connectprofiles ....... 87
Thermo Fisher
Connect Platform instrument profile roles andfunctions ......... 88
Link the instrument to your Thermo Fisher
Connect Platformaccount ........... 88
Link a local profile to a Thermo Fisher
Connectprofile ......................... 89
Unlink a Thermo Fisher
Connectaccount .................................... 91
If you link when you are signed in to theinstrument ............................ 91
Change the PIN for a Thermo Fisher
Connectprofile .......................... 92
Enable SAE mode (Administratoronly) ............................................ 92
Require instrument profile sign-in (Administratoronly) .............................. 93
Enable remote instrument monitoring (Administratoronly) ........................... 93
Update instrument software (Administratoronly) ................................... 94
Restore factorydefaults ......................................................... 94
Configure thenetwork .......................................................... 94
Set up a wiredconnection .................................................. 95
Set up a wirelessconnection ................................................ 95
Select a region for the Thermo Fisher
Connect Platform (Administratoronly) ......... 96
Manage the instrument name (Administratoronly) .................................. 96
Set the date and time (Administratoronly) ......................................... 96
Manage the signout timer (Administrator only) ..................................... 97
APPENDIXAInstall and connect the instrument to anetwork ................ 98
Workflow: Install and connect to anetwork ........................................ 98
Before youbegin ............................................................... 99
Unpack and install theinstrument ................................................ 99
Power on and follow the startup wizard .......................................... 100
Connect the computer to the instrument directly or to aLAN ....................... 100
Instrument and computerconnections ........................................... 101
Contents
QuantStudio
3 and 5 RealTime PCR Systems Installation, Use, and Maintenance Guide
7
Download and install the desktop software ....................................... 102
Computer requirements for the desktop software ............................. 102
Download the desktop software ............................................ 103
Install the desktop software ................................................ 103
Networking ................................................................... 103
Supported options for instrument and computerconnections .................. 103
Control and monitor networkedinstruments .................................. 105
Ethernet port overview .................................................... 105
Firewall ports that must beopen ............................................ 105
Networking guidelines and best practices ................................... 106
APPENDIXBTroubleshooting .................................................. 107
APPENDIXCParts and materials .............................................. 109
Kits, consumables, accessories, and reagents .................................... 109
Consumables (96well, 0.2mLformat) ........................................... 109
384-wellconsumables ......................................................... 110
Consumables (96-well, 0.1-mLformat) .......................................... 111
Accessories .................................................................. 112
General-use materials andconsumables ......................................... 112
APPENDIXDInstrument specification andlayout ............................ 113
Configured systemdimensions ................................................. 114
Instrument clearances ..................................................... 115
Electrical requirements ........................................................ 115
Environmental requirements .................................................... 116
Network requirements ......................................................... 116
APPENDIXESafety .............................................................. 118
Symbols on thisinstrument .................................................... 118
Conformitysymbols ...................................................... 120
Safety alerts on thisinstrument ................................................. 121
Location of safety labels on theinstrument .................................. 121
Safety information for instruments not manufactured by Thermo FisherScientific ..... 122
Instrumentsafety ............................................................. 122
General ................................................................. 122
Physical injury ............................................................ 123
Electricalsafety .......................................................... 124
Cleaning anddecontamination ............................................. 124
Contents
8
QuantStudio
3 and 5 RealTime PCR Systems Installation, Use, and Maintenance Guide
Safety and electromagnetic compatibility (EMC) standards ......................... 124
Safetycompliance ........................................................ 125
EMC .................................................................... 125
Environmentaldesign ..................................................... 126
Chemicalsafety .............................................................. 127
Biological hazardsafety ....................................................... 128
APPENDIXFDocumentation and support ..................................... 129
Relateddocumentation ........................................................ 129
Obtain information from the Help system ........................................ 129
Customer and technical support ................................................ 129
Limited product warranty ...................................................... 130
Index ..................................................................................... 131
Contents
QuantStudio
3 and 5 RealTime PCR Systems Installation, Use, and Maintenance Guide
9
About this guide
Purpose
This guide provides information about installing, using, and maintaining the QuantStudio
3 and 5
RealTime PCR Systems.
Note: For information and instructions on performing experiments on these systems, see QuantStudio
Design and Analysis Desktop Software User Guide (Pub. No. MAN0010408).
10
QuantStudio
3 and 5 RealTime PCR Systems Installation, Use, and Maintenance Guide
Product information
Network and password security requirements ............................................ 11
Perform internal validations before software upgrades ..................................... 11
Update the instrumentfirmware ........................................................ 12
Instrument hardwaredescription ....................................................... 12
Softwaredescription .................................................................. 16
Network connectionoptions ........................................................... 19
Types ofruns ........................................................................ 19
Network and password security requirements
Network configuration and security
The network configuration and security settings of your laboratory or facility (such as firewalls, anti-
virus software, network passwords) are the sole responsibility of your facility administrator, IT, and
security personnel. This product does not provide any network or security configuration files, utilities, or
instructions.
If external or network drives are connected to the software, it is the responsibility of your IT personnel
to ensure that such drives are configured and secured correctly to prevent data corruption or loss. It
is the responsibility of your facility administrator, IT, and security personnel to prevent the use of any
unsecured ports (such as USB, Ethernet) and ensure that the system security is maintained.
Password security
Thermo Fisher Scientific strongly recommends that you maintain unique passwords for all accounts in
use on this product. All passwords should be reset upon first sign in to the product. Change passwords
according to your organization's password policy.
It is the sole responsibility of your IT personnel to develop and enforce secure use of passwords.
Perform internal validations before software upgrades
IMPORTANT!
If you have a validated workflow, you must perform all internal validations as required by
your organization's standard operating procedures before performing a software upgrade.
1
QuantStudio
3 and 5 RealTime PCR Systems Installation, Use, and Maintenance Guide
11
Update the instrument firmware
If your instrument is not on version 1.6 or later of the firmware you must update the firmware to
access the features and updates described in Revision E.0 of the revision history. For information on
updating your firmware, see https://www.thermofisher.com/us/en/home/global/forms/life-science/
quantstudio-3-5-software.html.
Prior to updating your firmware, we recommend you perform a backup. For information on performing a
backup, see “Back up the instrument” on page12.
Back up the instrument
1.
In the home screen, touch Settings4Maintenance and Service4Backup/Restore.
2.
Touch Backup Instrument.
3.
Select a location to store the backup settings for your instrument.
4.
If backing up to a USB drive, insert a USB drive into the front-panel USB port.
5.
Enter a backup file name, then touch Done.
6.
Select the elements to backup, or leave them all selected.
7.
Touch Backup.
Instrument hardware description
Instrument overview
The QuantStudio
3 and 5 RealTime PCR Systems use fluorescence-based polymerase chain reaction
(PCR) reagents to perform:
Quantitative detection of target nucleic acid sequences (targets).
Qualitative detection of targets (endpoint analysis, genotyping experiments, and presence/absence
experiments).
Qualitative analysis of the PCR product (post-PCR melt curve analysis).
Chapter1Product information
Update the instrument firmware
1
12
QuantStudio
3 and 5 RealTime PCR Systems Installation, Use, and Maintenance Guide
The following fixed-block configurations are available:
QuantStudio
3 Real-Time PCR System QuantStudio
5 RealTime PCR System
96-Well VeriFlex
0.2-mL Block (4Color)
96-Well VeriFlex
0.1-mL Block (4Color)
96-Well VeriFlex
0.2-mL Block (6ColorDecoupled)
384-Well Block (5Color)
96–Well VeriFlex
0.1-mL Block (6ColorDecoupled)
The instrument can be run directly from the touchscreen to create and start experiments. To design
experiments or to analyze data, the instrument can be integrated with the QuantStudio
Design and
Analysis Software.
An optional barcode scanner and optional wireless adapter can be purchased separately.
Instrument filters and supported dyes
System dyes
The instruments use a coupled four-color, coupled five-color, or de-coupled six-color filter set that
supports the dyes shown in the following table and figure. For more information about available spectral
dye calibration kits, contact Support.
Peak
filter
Color
Filter wavelength (nm)
[1]
Factory-calibrated dyes Example customdyes
Excitation Emission
x1-m1 Blue 470 ± 15nm 520 ± 15nm
FAM
dye, SYBR
Green
dye
SYT09
x2-m2 Green 520 ± 10nm 558 ± 12nm VIC
dye
JOE
dye,
HEX
dye, TET
dye
[2]
x3-m3 Yellow 550 ± 10nm 587 ± 10nm
NED
dye, TAMRA
dye,
ABY
dye
Cy
®
3 dye
x4-m4 Orange 580 ± 10nm 623 ± 14nm ROX
dye, JUN
dye Texas Red
dye
x5-m5 Red 640 ± 10nm 682 ± 14nm
MUSTANG PURPLE
dye,
Cy
®
5
LIZ
dye
x6-m6
Deep-
Red
662 ± 10nm 711 ± 12nm None
[3]
Cy
®
5.5 dye
[1]
The central wavelengths are the optimized wavelengths.
[2]
The HEX
and TET
dyes from Thermo Fisher Scientific fall within the emission wavelength range of the system, therefore they can
be added and adapted for use on the instrument.
[3]
This filter set currently does not support any dyes supplied by Thermo Fisher Scientific.
Chapter1Product information
Instrument hardware description
1
QuantStudio
3 and 5 RealTime PCR Systems Installation, Use, and Maintenance Guide
13
Filters
Wavelength
(nm)
1
2
3
4
5
x1-m1 x2-m2
x3-m3
x4-m4
x5-m5 x6-m6
Emission
Spectra
1
x1-m1 — FAM
dye, SYBR
Green dye
2
x2-m2 — VIC
dye
3
x3-m3 — ABY
dye, NED
dye, Cy
®
3 dye, TAMRA
dye
4
x4-m4 — JUN
dye, ROX
dye, Texas Red
dye
5
x5-m5 — Cy
®
5 dye, MUSTANG PURPLE
dye
Custom dyes
The instrument can run assays designed with custom dyes.
Custom dyes include:
Dyes that are not supplied by Thermo Fisher Scientific.
Dyes or formulations of dyes that are not system dyes for the instrument.
To use a custom dye on the instrument, review the following requirements.
Calibrate the instrument for the custom dye (see “Calibrate custom dyes” on page65).
Ensure that the custom dye excites between 455 – 672 nm and emits between 505 – 723 nm.
Select a custom dye that does not overlap with other dyes used in the run (see the filter-
wavelength table in “System dyes” on page13).
The custom dye must be attached to the 5' end of a short DNA oligonucleotide consisting of the
first two bases of the probe sequence without a quencher at the 3'end.
Chapter1Product information
Instrument hardware description
1
14
QuantStudio
3 and 5 RealTime PCR Systems Installation, Use, and Maintenance Guide
About data collection
The instrument collects raw fluorescence data at dierent points during the PCR cycle, depending on
the type of run performed.
When you create an experiment template (EDT file) in the software, you can customize the optical filter
channels through which the instrument collects data. You can specify a filter channel set for all PCR
thermal protocols and, optionally, a dierent filter set for the melt curve stages.
Run type Experiment type Data collection point
Real-time
Standard curve
Relative standard
curve
Comparative C
T
(ΔΔC
T
)
Melt curve
During the thermal cycling protocol.
Typical timing is to collect data at each cycle of a PCR stage
or continuously during a melt stage.
Post-PCR
(endpoint)
Genotyping
Presence/ Absence
After thermal cycling is completed.
For Presence/Absence and Genotyping experiments,
data collection before the PCR cycle is optional, but
recommended.
(Optional) Before thermal cycling starts.
Collecting data during the run can confirm genotyping
results by viewing traces in allelic discrimination plots or
viewing genotyping calls at earlier cycles.
Blocks with VeriFlex
Zones
Applied Biosystems
VeriFlex
technology provides independent temperature zones that oer
enhanced functionality and precise control over your real-time PCR runs.
QuantStudio
3 System QuantStudio
5 System
Three programmable VeriFlex
Zones. Six programmable VeriFlex
Zones.
[1]
[1]
Only applicable for 96-well 0.2-mL and 0.1-mL blocks.
Chapter1
Product information
Instrument hardware description
1
QuantStudio
3 and 5 RealTime PCR Systems Installation, Use, and Maintenance Guide
15
These independent zones are ideal for real-time PCR optimization or the ability to run multiple
experiments in the same run. Unlike standard gradients which give a sigmoidal temperature curve
across the columns, blocks with VeriFlex
Zones help deliver accurate temperatures across every zone.
Parts of the instrument
1
Touchscreen – Controls the instrument.
2
USB port – For connection to an external network drive or external data storage device.
3
Instrument drawer – Contains sample plate.
The instrument includes three additional USB ports on the back of the instrument.
Note: The instrument recognizes only one external storage device at a time for data transfer.
Software description
Instrument, desktop, and cloud software features
The instrument software and the QuantStudio
Design and Analysis Software (desktop and Thermo
Fisher
Connect Platform) include the features described below.
Feature
Instrument Desktop
Thermo Fisher
Connect Platform
Use as guest (no sign in)
[1]
Create templates (unlocked or locked)
Edit unlocked templates
Edit locked templates (password assigned by template creator
required)
Load system or user-created templates (EDT file) to instrument
Chapter1Product information
Software description
1
16
QuantStudio
3 and 5 RealTime PCR Systems Installation, Use, and Maintenance Guide
(continued)
Feature Instrument Desktop
Thermo Fisher
Connect Platform
Change experiment settings in template (EDT file) loaded on
the instrument
Settings that can be changed in a locked template (no
password required):
Properties: All settings
Method: No changes allowed
Plate: Sample names
Settings that can be changed in an unlocked template:
Properties: All settings
Method: All settings
Plate: Sample names
Load plate in instrument
Start run
View real-time data during a run
View instrument status (running, calibration needed, and so on)
Analyze results
Set calibration reminders
Review exported calibration or RNaseP verification results
(only RNase P)
[1]
Login required if the Security, Audit, and e-Signature (SAE) module is enabled (QuantStudio
5 System only)
Folders, templates, experiments, and projects
Term
Definition Supported in
Folders in Load
Experiment
Location in which you can store templates (EDT files) on the
instrument:
My Instrument – Displayed if you are signed in
Public – Location in which all experiments run by guest users
are stored
USB – USB for manual transfer to and from a computer
Post Read – If you run an endpoint experiment, the pre-read
experiment is automatically saved in this folder for post-read
analysis
Instrument
Chapter1Product information
Software description
1
QuantStudio
3 and 5 RealTime PCR Systems Installation, Use, and Maintenance Guide
17
(continued)
Term
Definition Supported in
Experiment
Template File
(EDT)
Default settings for an instrument run, can be modified before
instrument run.
Two types:
Factory-provided, accessed from Open Template (instrument)
or Create Experiment (desktop or Thermo Fisher
Connect
Platform)
User-created, accessed from Load Experiment
Instrument
Desktop
Thermo Fisher
Connect Platform
Experiment Run
File (EDS)
Settings and data for a completed instrument run. Instrument
Desktop
Thermo Fisher
Connect Platform
Template and
experiment
components
Instrument Desktop
Thermo Fisher
Connect
Platform
Properties
Experiment file name
Reagent information
(Reagent barcode, Lot#)
Plate barcode
Data management (location
for auto-transfer of data)
Comments (Tags are not
used at this time)
Experiment file name
Plate barcode
User name
Instrument type
Block type
Experiment type
Chemistry (reagent information)
Run mode
Comments
(Thermo Fisher
Connect Platform only) Notifications
Method Thermal cycling conditions Thermal cycling conditions
Plate Sample names
You cannot edit targets/SNP
assays or tasks on the
instrument.
Define and assign samples, targets or SNP assays, and tasks in
the Quick Setup and Advanced Setup panes of the Plate tab.
Run Start and monitor a run in
progress
View: time remaining and
temperature, method, plots
Pause, resume, stop a run
Start and monitor a run in
progress
View: time remaining and
temperature, method, plots
Monitor a run in progress (link
to Thermo Fisher
Connect
Platform Instrument Details)
View: time remaining and
temperature, method, plots
Chapter1Product information
Software description
1
18
QuantStudio
3 and 5 RealTime PCR Systems Installation, Use, and Maintenance Guide
(continued)
Template and
experiment
components
Instrument Desktop
Thermo Fisher
Connect
Platform
Results Not applicable Review plots
Export Not applicable Export results
Third-party software
Before installing third-party software on the computer running QuantStudio
Design and Analysis
Software, confirm that the third-party software will not do the following:
Restrict Ethernet communication.
Interfere with instrument or computer operation.
Network connection options
You can connect an instrument to a network or computer in the configurations listed below. For specific
information on networking, see “Networking” on page103.
Direct connection
Local area network (LAN) connection
Thermo Fisher
Connect Platform
connection
wired wired or wireless wired or wireless
LAN
LAN
LAN
Types of runs
Purpose
Description
Standard curve run
Determines absolute
target quantity in
samples.
1.
The software measures amplification of the target in a standard dilution series and
in test samples.
2.
The software generates a standard curve using data from the standard dilution
series.
3.
The software uses the standard curve to interpolate the absolute quantity of target
in the test samples.
Chapter1Product information
Network connection options
1
QuantStudio
3 and 5 RealTime PCR Systems Installation, Use, and Maintenance Guide
19
Purpose
Description
Relative standard curve run
Determines relative target
quantity in samples.
1.
The software measures amplification of the target of interest and of an endogenous
control target in a standard dilution series, in a reference (calibrator) sample, and in
test samples.
The endogenous control is a target that is expressed equally in all samples;
examples of endogenous controls are β‐actin, GAPDH, and 18S ribosomal RNA.
The software can algorithmically incorporate multiple endogenous control targets in
relative
quantification calculations.
The reference sample is used as the basis for relative quantification results (or
1× sample). For example, in a study of drug eects on gene expression, an
untreated control is an appropriate reference sample.
2.
The software generates standard curves for the target of interest and the
endogenous control using data from the corresponding standard dilution series.
3.
The software uses the standard curves to interpolate the quantities of the target of
interest and the endogenous control in each sample. The target quantity in each
sample is then normalized to the sample's endogenous control quantity.
4.
To determine the relative quantity of the target in test samples, the software divides
the normalized target quantity in the sample by the normalized target quantity in
the reference sample.
Comparative C
t
(ΔΔC
t
) run
Determines relative target
quantity in samples.
1.
The software measures amplification of the target of interest and of an endogenous
control target in a reference (calibrator) sample and in test samples.
The endogenous control is a target that is expressed equally in all samples;
examples of endogenous controls are β‐actin, GAPDH, and 18S ribosomal RNA.
The software can algorithmically incorporate multiple endogenous control targets in
relative
quantification calculations.
The reference sample is used as the basis for relative quantification results (or
1× sample). For example, in a study of drug eects on gene expression, an
untreated control is an appropriate reference sample.
2.
The measurements for the target of interest are normalized to the endogenous
control.
3.
To determine the relative quantity of the target in test samples, the software
compares the normalized ΔC
q
(ΔC
t
or ΔC
rt
) for the sample to the normalized
ΔC
q
(ΔC
t
or ΔC
rt
) for the reference sample.
Chapter1Product information
Types of runs
1
20
QuantStudio
3 and 5 RealTime PCR Systems Installation, Use, and Maintenance Guide
Purpose
Description
Genotyping run
Detects single nucleotide
polymorphism (SNP)
variants of a target
nucleic acid sequence.
Genotyping runs use preformulated TaqMan
SNP Genotyping Assays that include the
following components:
Two sequence-specific primers for amplification of sequences containing the SNP
of interest
Two allele-specific TaqMan
probes for Allele 1 and Allele 2
1.
The software normalizes the fluorescence of the reporter dyes to the fluorescence
of the passive reference dye in each well.
2.
The software plots the normalized reporter dye signal of each sample well on
an Allelic Discrimination Plot, which contrasts the reporter dye intensities of the
allele-specific probes.
3.
The software algorithmically clusters the sample data, and assigns a genotype call
to the samples of each cluster according to its position on the plot.
Presence/absence run
Determines the presence
or absence of a target
nucleic acid sequence in
a sample.
The software calls the target present or absent based on an algorithmically determined
call threshold. (The call threshold is dierent from the C
t
threshold; the C
t
threshold is
not used to make calls.)
Melt curve run
Determines the melting
temperature (T
m
) of the
amplification products
of a PCR that used
intercalating dyes.
In the software, melt curve analysis is included in the default run method for any run
type that uses intercalating dyes.
1.
The software plots a melt curve based on the fluorescence of the dye with respect
to change in temperature.
2.
Using the melt curve, the software calculates the melting temperature (T
m
).
Chapter1Product information
Types of runs
1
QuantStudio
3 and 5 RealTime PCR Systems Installation, Use, and Maintenance Guide
21
Start, sign on, and configure the
instrument
Installation and instrumentverification .................................................. 22
Precautions foruse ................................................................... 23
Power on theinstrument .............................................................. 23
Parts of the home screen .............................................................. 24
Use the instrument without signingin ................................................... 26
Create a new instrumentprofile ........................................................ 26
Signin .............................................................................. 27
Signout ............................................................................. 27
Configure instrumentsettings .......................................................... 27
Installation and instrument verification
Before the first use of the instrument, complete the following tasks:
Install the instrument (see “Unpack and install the instrument” on page99).
Verify instrument performance (see “Perform instrument verification using RNase P plates” on
page59).
Note:
·
Instruments are factory calibrated, so no calibration is necessary at installation. However, we
recommend that you verify instrument performance before using the instrument.
·
Regular calibration and verification should be performed according to the calibration and
verification schedule (see “Calibration and verification schedule” on page49).
2
22
QuantStudio
3 and 5 RealTime PCR Systems Installation, Use, and Maintenance Guide
Precautions for use
CAUTION! PHYSICAL INJURY HAZARD. Do not remove the instrument cover. There are no
components inside the instrument that you can safely service yourself. If you suspect a problem,
contact technical support.
CAUTION! FIRE HAZARD. For continued protection against the risk of fire, replace fuses only with
listed and certified fuses of the same type and rating as those currently in the instrument.
CAUTION! PHYSICAL INJURY HAZARD. During instrument operation, the sample block
temperature can reach 100°C. Allow it to cool to room temperature before handling.
CAUTION! Before using a cleaning or decontamination method other than those recommended by
Thermo Fisher Scientific, confirm with Thermo Fisher Scientific that the proposed method will not
damage the instrument.
CAUTION! Use flat caps for tubes. Rounded caps can damage the heated cover.
Power on the instrument
1.
Touch anywhere on the touchscreen to determine if the instrument is in sleep mode. If the home
screen is displayed, the instrument is already powered on.
2.
If the home screen does not display, power on the
instrument by pressing the switch on the rear panel.
If left unattended, the instrument automatically enters sleep
mode (enabled by default) to conserve power.
Note: To customize the sleep mode setting, touch
Settings4Instrument Settings4Sleep Mode.
Chapter2Start, sign on, and configure the instrument
Precautions for use
2
QuantStudio
3 and 5 RealTime PCR Systems Installation, Use, and Maintenance Guide
23
Parts of the home screen
4
2
9
1
5
6
8 7
3
1
Avatar and Instrument name
2
Eject icon
3
Help icon
4
Status dial
5
Current user name; instrument block type
6
Settings button
7
Buttons for accessing EDT files
8
Connectivity icons
9
Sign In (or My Profile) button
Chapter2Start, sign on, and configure the instrument
Parts of the home screen
2
24
QuantStudio
3 and 5 RealTime PCR Systems Installation, Use, and Maintenance Guide
Table1Parts of the home screen
Element of the home
screen
Function
For more information,
see...
Avatar and Instrument
name
Set by the administrator to uniquely identify instrument. “Manage the instrument
name (Administrator
only)” on page96.
Eject icon Touch to open or close the instrument drawer.
Help icon Touch to launch the touchscreen Help system to access step-by-
step instructions.
Status dial
When the instrument is in use – Displays the sample block
temperature, the elapsed run time, and the run status.
You can swipe the dial to the left or touch to access
real-time views of the run.
When the instrument is not in use – Displays Set up run. You
can start a run by touching the status dial.
When the instrument is locked – Displays a lock icon ( )
within the status dial.
Current user name
and block type
Displays the user name of the current signed-in user and the
instrument block type.
Note: If no user is signed-in, the instrument defaults to the Guest
profile.
Settings button Touch Settings to configure, calibrate, or learn about the
instrument.
“Configure instrument
settings” on page27.
Buttons for accessing
experiment and
template files
Load Experiment – Touch to open a user-created EDT file from
a Thermo Fisher
Connect Platform account, USB, instrument
folder, or network drive.
“Run an experiment
from a saved file” on
page33.
Open Template – Touch to open a system EDT template file. “Create and run an
experiment from a
template” on page33.
Run Last – Touch to open the last EDT file run on the instrument.
If you are signed-in, the file opened will be the last file you
ran when signed-in.
If you are not signed-in, the file opened will be the last file the
guest profile ran.
“Repeat the last
instrument run” on
page34.
Connectivity icons
—The instrument is connected to a wired network.
—The instrument is wirelessly connected.
—A USB drive is plugged into the instrument.
—The instrument is linked to a Thermo Fisher
Connect
Platform account.
Chapter2Start, sign on, and configure the instrument
Parts of the home screen
2
QuantStudio
3 and 5 RealTime PCR Systems Installation, Use, and Maintenance Guide
25
Table 1 Parts of the home screen(continued)
Element of the home
screen
Function
For more information,
see...
Sign In button
(My Profile button
when a user is signed
in)
Touch Sign In to sign into an instrument profile and link to
a Thermo Fisher
Connect Platform account.
Touch My Profile to change instrument profile settings,
link to a Thermo Fisher
Connect Platform account, or lock
the instrument during a run.
“Sign in” on page27.
Use the instrument without signing in
If the instrument is configured by an Administrator to allow guest access ( Settings4Manage
Users4Sign In Required set to o), you can use the instrument without signing in.
The following limits are in place if you do not sign in to the instrument:
All actions are recorded to Guest user profile.
You have access only to the Public folder for selecting and storing experiments.
You cannot transfer data to Thermo Fisher
Connect Platform (only to USB or network drive).
Create a new instrument profile
1.
In the home screen, touch SignIn, then touch Get Started.
2.
Touch Name, enter a username, then touch Done.
3.
Touch PIN Code, enter a four-digit numerical password, then touch Enter.
Note: Touch the Show PIN checkbox to switch PIN display on or o.
4.
Touch Confirm PIN, then repeat step2.
5.
Touch Create profile.
6.
Sign in to the profile you just created.
Note: To enable access to Thermo Fisher
Connect Platform, see “Link an instrument profile to
Thermo Fisher
Connect” on page86.
Chapter2Start, sign on, and configure the instrument
Use the instrument without signing in
2
26
QuantStudio
3 and 5 RealTime PCR Systems Installation, Use, and Maintenance Guide
Sign in
Create an instrument profile before signing into the instrument. See “Create a new instrument profile”
on page26.
Note: An instrument profile is a user account specifically for the instrument. It is not related to any
other user account for the system or software.
1.
In the home screen, touch SignIn.
2.
Touch SignIn, then select your username.
3.
Enter your PIN, then touch Enter.
Note: To enable access to Thermo Fisher
Connect Platform, see “Link an instrument profile to
Thermo Fisher
Connect” on page86.
Sign out
1.
In the home screen, touch MyProfile.
2.
Touch SignOut.
Configure instrument settings
Touch Settings in the home screen to configure settings as needed.
Note: When SAE mode is enabled, the Settings button is not available if a user is not signed in.
Touch for step-by-step instructions on configuring settings.
Chapter2
Start, sign on, and configure the instrument
Sign in
2
QuantStudio
3 and 5 RealTime PCR Systems Installation, Use, and Maintenance Guide
27
Options Description
Instrument Settings
Instrument Name
(Administrator only)
Instrument name. If connected to a network, the instrument name must be unique.
(Optional) Add an avatar image for the instrument (JPG, PNG, or GIF) from a USB.
Sleep Mode Enable the instrument to enter a standby mode after a set length of inactivity.
Heated Cover
Temperature
Set the temperature of the heated cover for instrument operation and standby mode.
Enable heated cover to automatically turn o during standby mode.
Network Drive Specify a default network location for the signed-in user.
Insert Plate Reminder Enable a reminder to insert a plate before starting a run from the instrument.
OEM Connection Only
(Administrator only)
Required for API access to the instrument. When enabled, the desktop cannot connect
to the instrument.
API access to the instrument is exclusive to authorized OEM partners.
Cloud Region
(Administrator only)
Specify the regional server location to access Thermo Fisher
Connect Platform.
Note: Once you set a region for the cloud platform, restore factory defaults to change
the region.
Date/Time Set time zone and date and time formats.
Network Connection Set wired or wireless network connection for the instrument.
Restore Factory Defaults Restore the instrument to the factory settings.
IMPORTANT! Back up the instrument before restoring factory defaults (see “Back up
or restore the instrument” on page74).
Note: When you restore an instrument to its factory defaults:
·
User profiles and files stored on the instrument are deleted, including all user-
created EDT and EDS
files and any custom dye and custom melt calibrations.
·
System templates and factory calibrations remain on the instrument.
SAE Mode
(Administrator only)
(QuantStudio
5 RealTime
PCR Instrument only)
When enabled the following items apply:
The home screen displays SAE mode.
Runs must be started from the desktop software. The instrument locks out access
to load experiments, open templates, repeat the last run, start a run, change the
instrument name, and change the date/time.
Only an administrator can perform calibrations.
A user can monitor a run and view the method from the instrument touchscreen.
The Settings button is not available if a user is not signed in.
Chapter2Start, sign on, and configure the instrument
Configure instrument settings
2
28
QuantStudio
3 and 5 RealTime PCR Systems Installation, Use, and Maintenance Guide
Options Description
About Instrument
About Instrument Displays the instrument Model Name, Block Serial Number, and Firmware Version.
License Agreement Displays the End User Software License Agreement and the Limited Product Warranty.
You can export the License Agreement to a USB drive.
Notifications
Enable home screen notifications of instrument errors and software updates. This
function is not related to the Notifications function in the cloud software.
The number of new, unviewed notifications displays over Settings in the home screen.
Maintenance and Service
Software Update
(Administrator only)
Update the instrument software.
Monitoring
(Administrator only)
Enable:
Remote Monitoring Service to automatically send critical device statistics to
Thermo Fisher Scientific (does not monitor or send data).
Thermo Fisher Scientific Cloud Monitor to allow real-time monitoring of instrument
runs from a Cloud account.
Instrument Statistics Displays instrument usage information: Block Cycle Count and LED Life.
Calibrations
Perform calibrations
ROI and Uniformity
Dye
Custom (including Background calibration)
View calibration history and set calibration reminders in History and Reminders
RNaseP Verification Perform instrument verification using an RNase P plate.
Self Verification Test Check the instrument hardware functions.
Log View and export Instrument Log.
Backup / Restore
Backup Instrument
Restore a Backup (Administrator only)
Ship Prep Mode Place the instrument in a safe state for shipping, moving, or long-term storage.
Run History
Displays the experiments run on the instrument and whether the data was transferred.
Note: Experiments run while a user is signed-in can only be viewed or transferred by
that user or an administrator.
Touch an experiment to view experiment run details and to transfer or delete EDS file.
Chapter2Start, sign on, and configure the instrument
Configure instrument settings
2
QuantStudio
3 and 5 RealTime PCR Systems Installation, Use, and Maintenance Guide
29
Options Description
Manage Users
Sign In Required
(Administrator only)
Enable the restriction of instrument use to only signed-in users (disables Guest profile
use).
Sign Out Timer
(Administrator only)
Set the duration of inactivity before a user is automatically signed out.
Manage Profiles
(Administrator only)
Instrument
Cloud
All Profiles
Displays the profile information for the instrument and the associated Thermo Fisher
Connect Platform account.
Data Management
User defined or System
defined
(Administrator only)
When data management is system-defined, the destination to save the data cannot be
edited by an individual user.
When data management is system-defined, the Thermo Fisher
Connect Platform
cannot be selected as a location to save the data. The Thermo Fisher
Connect
Platform is specific to each user and cannot be system-defined.
When data management is user-defined, each individual user can select the
destination to save the data.
USB and Network
(Administrator only)
Select the default destination to save the data.
Automatically transfer run
data (EDS files)
Select the Automatically transfer run data (EDS files) checkbox to automatically
transfer EDS files after the run. The Automatically transfer run data (EDS files)
checkbox is available for both the USB drive destination and the network drive
destination.
Chapter2Start, sign on, and configure the instrument
Configure instrument settings
2
30
QuantStudio
3 and 5 RealTime PCR Systems Installation, Use, and Maintenance Guide
Create and run experiments on the
instrument
Workflow ............................................................................ 32
Options for running an experiment ...................................................... 33
Edit an experiment before starting arun ................................................. 35
Options for loading MicroAmp
consumables with the QuantStudio
3 and 5 RealTime
PCR Systems ........................................................................ 40
Load and unload the plate ............................................................ 40
Prepare samples using MicroAmp
tubes/tube strips with separate capstrips ............... 43
View, pause, or stop arun ............................................................. 45
Transfer, view, or manage files and results ............................................... 46
3
QuantStudio
3 and 5 RealTime PCR Systems Installation, Use, and Maintenance Guide
31
Workflow
Start the instrument
(Optional) Sign in to your instrument profile
(Optional) Link to your Thermo Fisher
Connect Platform account
Load an existing experiment template (EDT file) or
Create an experiment template (EDT file)
Modify experiment settings as needed
Load the plate in the instrument
Start the run from the instrument or the desktop software
(Optional) Monitor the run from Thermo Fisher
Connect Platform
Transfer results (EDS file) to the desktop software or Thermo Fisher
Connect Platform, then analyze
Chapter3Create and run experiments on the instrument
Workflow
3
32
QuantStudio
3 and 5 RealTime PCR Systems Installation, Use, and Maintenance Guide
Options for running an experiment
Create and run an experiment from a template
If an Administrator has set data management as system-defined, you cannot select the location to save
the data. See “Configure instrument settings” on page27.
1.
In the home screen, touch Open Template.
2.
(Optional) Touch a category in the left column.
3.
Touch the file name.
4.
(Optional) Enter or edit template properties, including Run File Name (EDS file name), Plate
Barcode, Reagent Information, and Data Destination (see “Enter or edit template properties”).
5.
(Optional) Edit the run method (see “Edit the run method” on page36).
Add, remove, or edit a step, stage, melt curve, or data collection point.
Adjust the heated cover temperature, sample volume, or number of cycles.
Configure VeriFlex
Zones, ramp rate, and pause settings.
6.
(Optional) Define plate wells with sample names and view Well ID, Targets, or Dyes (see “Define,
assign, and view well details”).
7.
Load a plate into the instrument (see “Load and unload the plate” on page40).
8.
Touch Start Run.
When prompted, confirm that you inserted a plate.
Note: To disable this reminder, select Do not show again or select Settings4Insert Plate
Reminder in the home screen.
Run an experiment from a saved file
If an Administrator has set data management as system-defined, you cannot select the location to save
the data. See “Configure instrument settings” on page27.
1.
In the home screen, touch Load Experiment.
2.
Touch the appropriate icon to navigate to your file location.
For files saved to the guest profile, touch
MyInstrument4Public.
For pre and post read files, touch
MyInstrument4Post Read.
3.
Touch the file name.
4.
(Optional) “Manage templates (EDT files)” in either USB or
MyInstrument.
Chapter3
Create and run experiments on the instrument
Options for running an experiment
3
QuantStudio
3 and 5 RealTime PCR Systems Installation, Use, and Maintenance Guide
33
5.
(Optional) Enter or edit template properties, including Run File Name (EDS file name), Plate
Barcode, Reagent Information, and Data Destination (see “Enter or edit template properties”).
6.
(Optional) Edit the run method (see “Edit the run method” on page36).
Add, remove, or edit a step, stage, melt curve, or data collection point.
Adjust the heated cover temperature, sample volume, or number of cycles.
Configure VeriFlex
Zones, ramp rate, and pause settings.
7.
(Optional) Define plate wells with sample names and view Well ID, Targets, or Dyes (see “Define,
assign, and view well details”).
8.
Load a plate into the instrument (see “Load and unload the plate” on page40).
9.
Touch Start Run.
When prompted, confirm that you inserted a plate.
Note: To disable this reminder, select Do not show again or select Settings4Insert Plate
Reminder in the home screen.
Repeat the last instrument run
This feature applies only to runs started from the instrument and is not available for runs started from
the desktop software. If you are signed-in, this feature applies to the last run from your instrument
profile.
If an Administrator has set data management as system-defined, you cannot select the location to save
the data. See “Configure instrument settings” on page27.
1.
In the home screen, touch Run Last.
2.
(Optional) Enter or edit template properties, including Run File Name (EDS file name), Plate
Barcode, Reagent Information, and Data Destination (see “Enter or edit template properties”).
3.
(Optional) Edit the run method (see “Edit the run method” on page36).
Add, remove, or edit a step, stage, melt curve, or data collection point.
Adjust the heated cover temperature, sample volume, or number of cycles.
Configure VeriFlex
Zones, ramp rate, and pause settings.
4.
(Optional) Define plate wells with sample names and view Well ID, Targets, or Dyes (see “Define,
assign, and view well details”).
5.
Load a plate into the instrument (see “Load and unload the plate” on page40).
6.
Touch Start Run.
When prompted, confirm that you inserted a plate.
Note: To disable this reminder, select Do not show again or select Settings4Insert Plate
Reminder in the home screen.
Chapter3
Create and run experiments on the instrument
Options for running an experiment
3
34
QuantStudio
3 and 5 RealTime PCR Systems Installation, Use, and Maintenance Guide
Edit an experiment before starting a run
Enter or edit template properties
Access a template (EDT file). For more information, see the following sections:
“Create and run an experiment from a template” on page33.
“Run an experiment from a saved file” on page33.
“Repeat the last instrument run” on page34.
If an Administrator has set data management as system-defined, you cannot select the location to save
the data. See “Configure instrument settings” on page27.
In the Properties screen, touch Edit to enter or edit template properties.
Edit the file name for the EDS file for the run.
a.
Touch the Run File Name field.
b.
Enter the name for the EDS file, then touch Done.
Enter a plate barcode.
a.
Touch the Plate Barcode field.
b.
Enter or scan the plate barcode, then touch Done.
Record reagents and their expiration dates.
a.
Touch Reagent Information.
b.
Touch Add, or touch an existing reagent, then touch Edit or Delete.
c.
Touch the Name, Type, Lot #, Reagent Barcode, Part #, or Expiration Date field to enter
individual reagent information.
d.
Touch Done.
Automatically transfer run data (EDS file) when an instrument run ends.
a.
Touch Data Destination.
b.
Touch the appropriate icon to select a data destination.
c.
Under the desired data destination, select Automatically transfer experiment.
d.
Touch Done.
Enter a comment.
a.
Touch Comments.
b.
Enter text, then touch Done.
Chapter3
Create and run experiments on the instrument
Edit an experiment before starting a run
3
QuantStudio
3 and 5 RealTime PCR Systems Installation, Use, and Maintenance Guide
35
Scan a barcode using the optional barcode scanner
The instrument is compatible with an optional Handheld Barcode Scanner (Cat. No.4488442,
purchased separately). The barcode scanner reads Code 128 (alphanumeric), which supports
128 ASCII character barcodes.
1.
Click the Barcode field.
2.
Hold the scanner 20–30 cm away from a plate or container label and aim at the center of the
barcode, then press the trigger.
3.
Slowly move the scanning beam across the barcode until the scanner emits a high-pitched tone.
When the scanner scans a barcode, it automatically transmits the following information:
Transmits the alphanumeric equivalent of the barcode to the barcode field.
Transmits other reagent information (Lot #, Part #, Expiration Date, etc.)
For more information about the hand-held barcode scanner, see the user documentation provided with
the barcode scanner.
Edit the run method
Access a template (EDT file):
“Create and run an experiment from a template” on page33.
“Run an experiment from a saved file” on page33.
“Repeat the last instrument run” on page34.
For an overview of the method as it is graphically represented on the touchscreen, see “Method
elements” on page37.
In the Method screen:
1.
Touch Edit.
2.
Touch a field, enter changes, then touch Enter.
Note: Touch–drag to quickly increase or decrease a step temperature.
3.
Touch Manage Steps to add or remove a step, stage, melt curve, or data collection point.
Note: The Manage Steps screen also provides access to configuring VeriFlex
Zones, Ramp
Rates, and Add Pause settings (see “Configure VeriFlex
Zones, ramp rates, and pause settings”
on page39).
Chapter3Create and run experiments on the instrument
Edit an experiment before starting a run
3
36
QuantStudio
3 and 5 RealTime PCR Systems Installation, Use, and Maintenance Guide
Method elements
2
3
1
4
5
9
8
7
6
1
Method tab
2
Heated cover temperature
3
Reaction volume
4
Stage of thermal protocol
5
Step within a stage
6
Number of cycles for the stage
7
Temperature for the step
8
Hold time for the step
9
Data collection point
Manage steps, stages, melt curves, and data collection points
Add or remove a step, stage, melt curve, or data collection point using the Manage Steps option.
You can also configure ramp rates and pause settings from the Manage Steps screen (see “Configure
VeriFlex
Zones, ramp rates, and pause settings” on page39).
In the Method screen, touch Edit4Manage Steps.
Option
Procedure
Add a step
1.
Touch Add/Remove steps4Add steps.
2.
Touch on the left or right border of a step to add a step before or after the
step, respectively.
3.
Enter parameters for the new step, then touch Enter.
4.
Touch Done.
Remove a step
1.
Touch Add/Remove steps4Remove steps.
2.
Touch on the step to be removed, then touch Done.
Chapter3Create and run experiments on the instrument
Edit an experiment before starting a run
3
QuantStudio
3 and 5 RealTime PCR Systems Installation, Use, and Maintenance Guide
37
(continued)
Option Procedure
Add a stage
1.
Touch Add/Remove stages4Add stages.
2.
Touch on the left or right border of a stage to add a stage before or after
the stage, respectively.
3.
Edit parameters of the new stage in the Method screen. See “Edit the run
method” on page36.
4.
Touch Done.
Remove a stage
1.
Touch Add/Remove stages4Remove stages.
2.
Touch on the stage to be removed, then touch Done.
Add a melt curve
1.
Touch Melt curves4Add melt curve.
2.
Touch on the left or right border of a step to add a melt curve before or
after the step, respectively.
3.
Select Continuous or Step and hold.
4.
Touch melt curve parameters to edit, then touch Done.
5.
Touch Done.
Note: Depending on the experiment type, there can be restrictions on the addition
or placement of melt curves.
Remove a melt
curve
1.
Touch Remove melt curve.
2.
Touch on the melt curve to be removed, then touch Done.
Add or remove
data collection
points
1.
Touch Data collection location.
2.
Touch to switch data collection on or o.
Chapter3Create and run experiments on the instrument
Edit an experiment before starting a run
3
38
QuantStudio
3 and 5 RealTime PCR Systems Installation, Use, and Maintenance Guide
Configure VeriFlex
Zones, ramp rates, and pause settings
In the Method screen, touch Edit4Manage Steps4Advanced Options.
Option Procedure
Use VeriFlex
Zones (96well
blocks only)
1.
Touch VeriFlex
Zones.
2.
Touch on the step to apply VeriFlex
Zones.
3.
Touch each zone to edit the temperature, then touch Enter.
The background colors of the VeriFlex
Zones change with the temperature
edit. Zones with higher temperatures display a background brighter than
those with lower temperatures.
4.
(Optional) Apply additional VeriFlex
Zones.
5.
Touch Done.
Edit ramp rates
1.
Touch Ramp Rates.
2.
Touch the ramp rates fields.
3.
Touch Enter.
4.
(Optional) Edit additional steps, then touch Done.
Add a pause into
the method
1.
Touch Add Pause.
2.
Touch in a stage.
3.
Enter the pause temperature, and the cycle after which the pause will occur.
4.
Touch Enter.
The pause is represented by a P in the corner of the stage.
5.
Touch Done.
Define, assign, and view well details
1.
Access a template (EDT file). For more information, see the following sections.
“Create and run an experiment from a template” on page33.
“Run an experiment from a saved file” on page33.
“Repeat the last instrument run” on page34.
Chapter3Create and run experiments on the instrument
Edit an experiment before starting a run
3
QuantStudio
3 and 5 RealTime PCR Systems Installation, Use, and Maintenance Guide
39
2.
In the Plate screen, define, assign, and view well details.
Option Procedure
Define and assign samples in
the plate layout view
1.
Touch .
2.
Touch Manage or touch an individual well.
3.
Touch the Samples subtab.
4.
Select one or more wells, then touch Edit.
5.
Enter sample names for the selected wells, then touch Done.
Define and assign samples in
the well table view
1.
Touch .
2.
Touch Edit to edit sample names for individual wells.
3.
Touch a sample name field, enter a new name, then touch Done.
View target information
1.
Touch .
2.
Touch Manage or touch an individual well.
3.
Touch the Targets subtab.
4.
Touch Details to view target information, then touch Done.
3.
Touch Done to return to the Plate tab.
Options for loading MicroAmp
consumables with the
QuantStudio
3 and 5 RealTime PCR Systems
(Recommended) If using MicroAmp
reaction plates, proceed to “Load and unload the plate” on
page40.
If using MicroAmp
tubes/tube strips, proceed to “Prepare samples using MicroAmp
tubes/tube strips
with separate cap strips” on page43.
Load and unload the plate
If a compression pad is not used, see “Load and unload a plate in the instrument” on page40.
A compression pad is needed for some applications with a 96-well 0.2-mL MicroAmp plate.
See your product documentation to confirm if the use of a compression pad is required.
To load and unload the plate, see “(If applicable) load and unload a plate in the instrument with a
compression pad” on page41.
Load and unload a plate in the instrument
1.
Load the plate.
a.
Touch to eject the instrument drawer.
Chapter3
Create and run experiments on the instrument
Options for loading MicroAmp
consumables with the QuantStudio
3 and 5 RealTime PCR Systems
3
40
QuantStudio
3 and 5 RealTime PCR Systems Installation, Use, and Maintenance Guide
b.
Load the plate onto the plate adaptor so that the following criteria are met.
Well A1 of the plate is positioned in the top-left corner of the plate adapter.
The barcode faces the front of the instrument.
IMPORTANT! The instrument should be loaded and unloaded by trained operators who have
been warned of the moving parts hazard.
Note: (For 96-well 0.2-mL blocks only) Do not remove the black plate adapter before loading
a plate or tube strips. If used, tube strips can fit loosely in the adapter, but the heated cover
will apply the appropriate pressure to seat the tube strips securely in the adapter.
Note: The 384-well and 96-well Fast (0.1-mL) block configurations do not require a plate
adapter.
c.
Touch to close the instrument drawer.
2.
When the run ends, unload the plate.
a.
Touch to eject the instrument drawer.
b.
Remove the plate.
c.
Touch to close the instrument drawer.
CAUTION!
PHYSICAL INJURY HAZARD. During instrument operation, the plate temperature
can reach 100°C. Allow it to cool to room temperature before handling.
Note: If the instrument does not eject the plate, contact Support.
(If applicable)
load and unload a plate in the instrument with a compression
pad
Some applications use a compression pad for a 96-well 0.2-mL MicroAmp plate.
1.
Load the plate.
a.
Touch to eject the instrument drawer.
Chapter3
Create and run experiments on the instrument
Load and unload the plate
3
QuantStudio
3 and 5 RealTime PCR Systems Installation, Use, and Maintenance Guide
41
b.
Load the plate onto the plate adaptor so that the following criteria are met.
Position well A1 of the plate in the top-left corner of the plate adapter.
Ensure that the barcode faces the front of the instrument.
IMPORTANT! The instrument should be used by trained operators who have been warned of
the moving parts hazard.
Note: (For 96-well 0.2-mL blocks only) Do not remove the black plate adapter before loading
a plate or tube strips. If used, tube strips can fit loosely in the adapter, but the heated cover
will apply the appropriate pressure to seat the tube strips securely in the adapter.
c.
Before loading the plate in the drawer, place the MicroAmp
Optical Film Compression Pad
with the gray side down on the surface of the plate to help ensure a proper seal between the
instrument and the adhesive film.
IMPORTANT!
Be careful to place the compression pad with the brown side up and the
gray side down, centered on top of the plate. Ensure that the compression pad is free from
wrinkles and signs of deterioration prior to use.
d.
Touch to close the instrument drawer.
2.
When the run ends, unload the plate.
a.
Touch to eject the instrument drawer.
b.
Remove the plate.
Chapter3
Create and run experiments on the instrument
Load and unload the plate
3
42
QuantStudio
3 and 5 RealTime PCR Systems Installation, Use, and Maintenance Guide
c.
Touch to close the instrument drawer.
d.
Remove the MicroAmp
Optical Film Compression Pad from the plate and store it inside the
pack.
IMPORTANT!
·
If the compression pad becomes stuck inside the instrument, call service to inspect the
heated cover.
·
Between each use, place the pad back in the pouch so that it does not dry out.
·
Each compression pad may be used up to 20 times before discarding. Do not use more
than 20 times.
CAUTION! PHYSICAL INJURY HAZARD. During instrument operation, the plate temperature
can reach 100°C. Allow it to cool to room temperature before handling.
Note: If the instrument does not eject the plate, contact Support.
Prepare samples using MicroAmp
tubes/tube strips with
separate cap strips
CAUTION!
Use flat caps for 0.2-mL tubes and 0.1-mL tubes. Rounded caps can damage the heated
cover.
The following procedure describes how to properly load and seal MicroAmp tube strips using cap strips
and the MicroAmp 96-well tray/retainer set. For a visual demonstration, watch the video "How to use
adapter tray retainers" by visiting PCR/qPCR Plastics and Seals Education at thermofisher.com.
The MicroAmp
96-Well Tray/Retainer Set for Veriti
Systems (Cat.No. 4381850) for 0.2mL format has
a blue tray and retainer.
The MicroAmp
96-Well Tray for VeriFlex
Block (Cat.No. 4379983) for 0.1mL format has a black tray
and retainer.
1.
Separate the blue or black tray and retainer by squeezing the release catch as indicated in the
graphic.
1
2
3
1
1
Release catch
2
MicroAmp
96-Well Retainer
3
MicroAmp
96-Well Tray
Chapter3Create and run experiments on the instrument
Prepare samples using MicroAmp
tubes/tube strips with separate cap strips
3
QuantStudio
3 and 5 RealTime PCR Systems Installation, Use, and Maintenance Guide
43
2.
Place the blue or black tray on the 96-well base.
3.
Load the tube strips on the tray.
4.
Pipette the reaction mixture into the tubes.
5.
Place the blue or black retainer over the tubes and snap the retainer into the tray.
6.
Seal the tube strip using a MicroAmp
cap strip.
7.
Remove the blue or black tray/retainer assembly containing the sealed tube strips from the
96-well base and place the assembly on the instrument block with the black adapter on the block.
Note: Ensure the tubes and blue or black tray and retainer fit properly into the wells on the block.
The tube, tray, and retainer assembly may feel loose and demonstrate some movement within the
block, which is standard. The appropriate amount of pressure will be applied when the heated
cover is lowered onto it.
1
2
4
3
5
1
MicroAmp
8-Cap strip
2
MicroAmp
96-Well Retainer
3
MicroAmp
8-Tube Strip (0.2-mL) or MicroAmp
Reaction Tube without Cap (0.2-mL)
4
MicroAmp
96-Well Tray
5
MicroAmp
Splash Free 96-Well Base
Chapter3Create and run experiments on the instrument
Prepare samples using MicroAmp
tubes/tube strips with separate cap strips
3
44
QuantStudio
3 and 5 RealTime PCR Systems Installation, Use, and Maintenance Guide
View, pause, or stop a run
Do any of the following actions during an instrument run.
View real-time data and plots (see “View real-time data and plots on the instrument touchscreen”
on page45).
Adjust the display of real-time plots (see “Adjust the display of real-time plots on the instrument
touchscreen” on page46).
Touch or swipe left once to view realtime run method information or to edit the number of
cycles.
Touch or swipe left twice to view realtime data and plots.
Pause or stop an instrument run (see “Pause or stop an instrument run” on page45).
View real-time data and plots on the instrument touchscreen
1.
In the instrument home screen, during an instrument run, touch or swipe left twice.
2.
Touch Well details.
3.
Touch Samples, Targets, or Tasks to select a graphical representation of each selection.
4.
Touch Close to return to the home screen.
Pause or stop an instrument run
In the home screen, during an instrument run:
1.
Touch or swipe left once to view realtime run method information.
2.
Stop or pause the run.
Touch Stop Run.
Touch Pause, then enter a pause temperature.
CAUTION!
PHYSICAL INJURY HAZARD. During instrument operation, the plate temperature
can reach 100°C. If you want to access the plate during a run pause, enter room temperature as
the pause temperature and allow the plate to cool to room temperature before handling.
3.
(Optional) After pausing a run, touch Edit to change the number of cycles.
Chapter3Create and run experiments on the instrument
View, pause, or stop a run
3
QuantStudio
3 and 5 RealTime PCR Systems Installation, Use, and Maintenance Guide
45
Adjust the display of real-time plots on the instrument touchscreen
1.
In the instrument home screen, during an instrument run, touch or swipe left twice to view
real-time data and plots.
2.
Touch Zoom.
3.
Touch or to zoom in or out.
4.
Touch the arrows to pan left, right, up, or down on the graph.
5.
Touch Close to return to the default view.
Lock the touchscreen during a run
After you have started a run, you can lock the touchscreen so that other users cannot interfere with
instrument operation.
Note: You must be signed-in to use this feature.
1.
Lock the touchscreen.
a.
In the home screen, touch My Profile.
b.
Touch Lock Screen, then touch Lock.
2.
Unlock the touchscreen.
a.
Touch anywhere on the touchscreen.
b.
Touch the PIN Code field, then enter your PIN.
Note: The touchscreen automatically unlocks when the run is complete.
Transfer, view, or manage files and results
Transfer EDS files from the instrument home screen
If an Administrator has set data management as system-defined, you cannot select the location to save
the data. See “Configure instrument settings” on page27.
1.
In the home screen, when a run ends, touch Transfer File.
2.
Select the data destination for the EDS file.
3.
Navigate to and select a folder.
4.
Touch OK.
5.
Touch Transfer.
Chapter3
Create and run experiments on the instrument
Transfer, view, or manage files and results
3
46
QuantStudio
3 and 5 RealTime PCR Systems Installation, Use, and Maintenance Guide
Note: Touch Settings4Run History to transfer EDS files at any time.
View run history and delete or transfer files from the instrument
If an Administrator has set data management as system-defined, you cannot select the location to save
the data. See “Configure instrument settings” on page27.
When SAE mode is enabled, only an Administrator can delete a run.
In the home screen, touch Settings4Run History.
Touch an individual run record to view its details, then complete one of the following actions:
Touch Delete to delete the run record.
Touch Transfer to export the run data.
Touch Manage to select multiple run records for simultaneous viewing, deletion, or transfer.
Note:
·
Guests (users not signed-in) can only view guest run records.
·
Users signed into their instrument profiles can also view their own run records.
·
Administrators can view all run records.
Note: The instrument automatically resumes interrupted runs due to power outages within an hour,
even without software connection. If network issues disrupt the connection during a desktop-initiated
run, the run completes on the instrument with data saved in its history. For incomplete desktop-initiated
runs, data must be manually transferred to a USB or network drive. Data stays in the internal memory
on the instrument until manually deleted or reset.
Manage templates (EDT files)
This feature applies to templates (EDT files) on a USB or in
My Instrument. To manage run
results files (EDS files), see “Transfer EDS files from the instrument home screen” on page46.
If an Administrator has set data management as system-defined, you cannot select the location to save
the data. See “Configure instrument settings” on page27.
1.
In the home screen, touch Load Experiment.
2.
Touch USB or
My Instrument.
3.
Touch Manage Files.
4.
(Optional) Navigate through the My Instrument, Public, USB, and Post Read folders to access an
EDT file. Folder availability depends on your sign-in status.
5.
Touch one or more files to select for managing.
Chapter3
Create and run experiments on the instrument
Transfer, view, or manage files and results
3
QuantStudio
3 and 5 RealTime PCR Systems Installation, Use, and Maintenance Guide
47
6.
Delete or copy the files.
Touch Delete Files, then confirm deletion.
Touch Copy Files, select a file destination, then touch Paste Files.
7.
Touch Done.
Chapter3Create and run experiments on the instrument
Transfer, view, or manage files and results
3
48
QuantStudio
3 and 5 RealTime PCR Systems Installation, Use, and Maintenance Guide
Calibrate and verify instrument
performance
Calibration and verificationschedule .................................................... 49
Calibrationdescriptions ............................................................... 50
View the calibration status and set reminders ............................................ 51
Perform ROI/uniformity, background, and dye calibrations ................................. 52
Perform instrument verification using RNase P plates ..................................... 59
Calibrate customdyes ................................................................ 65
Calibrate for a custom melt curve experiment ............................................ 72
Calibration and verification schedule
The instrument is factory-calibrated and does not require calibration at installation. To help ensure
optimal performance, perform calibrations at the recommended frequency.
Note: After instrument installation, we recommend performing instrument verification using the
provided RNase P plate.
IMPORTANT!
Perform calibrations and instrument runs under the environmental conditions that
are specified in “Environmental requirements” on page116. Exposure to extreme temperatures can
adversely aect the instrument performance and shorten the life span of the instrument components.
To set the calibration frequency for the instrument, touch Settings4Maintenance and
Service4Calibrations4History and Reminders4Edit4Exp interval field.
Calibration
Recommended frequency
ROI/Uniformity
Every two years (recommended)
Always perform new Background and Dye calibrations after an ROI/Uniformity
calibration.
Note: Performing an ROI/Uniformity calibration invalidates all other calibrations.
4
QuantStudio
3 and 5 RealTime PCR Systems Installation, Use, and Maintenance Guide
49
(continued)
Calibration Recommended frequency
Background
Every two years (recommended)
Background calibration can also be performed, as needed:
To check for contamination (depends on usage and laboratory conditions).
To obtain the most accurate data for the removal of background
fluorescence.
Note: Performing a Background calibration does not invalidate any other calibration.
Dye
Every two years (recommended)
During a Dye calibration, only the dyes on the given spectral calibration plate are
calibrated.
Note: Performing a Dye calibration for a given dye plate does not invalidate any other
calibration.
RNaseP
instrument
verification
After installing or moving the instrument
After performing instrument calibrations (recommended)
As needed to confirm instrument performance
Note: To prepare custom dye plates and to perform custom calibrations, see “Calibrate custom dyes”
on page65.
Calibration descriptions
Calibration description and purpose
Pass Criteria
ROI/Uniformity
The software captures images for each optical filter.
The software uses calibration data to map the increase in fluorescence
to the plate wells during subsequent runs and to evaluate well-to-well
consistency of the signals.
The image for each filter
distinguishes all wells of the
plate.
Each well in the image is
distinct.
Background
The software captures background images for each optical filter in the
absence of sample and reagent, and it checks that the fluorescence from
each well is below a fluorescence threshold.
The software uses calibration data to remove background fluorescence
during a run.
Note: You can also run this calibration to determine if contamination is
related to the sample block or the plate.
The plate images for all
filters are free of abnormal
fluorescence.
Chapter4Calibrate and verify instrument performance
Calibration descriptions
4
50
QuantStudio
3 and 5 RealTime PCR Systems Installation, Use, and Maintenance Guide
(continued)
Calibration description and purpose
Pass Criteria
Dye
The software extracts a spectral profile for each dye standard, then
produces a set of spectral profiles plotted as fluorescence vs filter.
The software uses calibration data to characterize and distinguish the
individual contribution of each dye in the total fluorescence signals collected
by the instrument.
Dye spectra peak within the
same filter as their group.
View the calibration status and set reminders
View calibration status and set reminders in the instrument
If an Administrator has set data management as system-defined, you cannot select the location to save
the data. See “Configure instrument settings” on page27.
1.
In the home screen, touch Settings4Maintenance and Service4Calibrations4History and
Reminders.
2.
In the Calibration Reminders screen, view the status of each calibration type.
3.
(Optional) Touch a calibration row to view the history of that specific calibration type, then touch
Done.
4.
Touch Edit to set the calibration reminder settings. For each calibration type:
a.
Slide the control On to enable the calibration reminder.
b.
Edit the Exp interval and Remind me fields.
c.
Touch Save.
5.
(Optional) To transfer the calibration report, touch Export then follow the directions on the screen.
6.
Touch Done.
View calibration status and set reminders in Thermo Fisher
Connect
Platform
Note: The calibration reminders feature requires a connection between the instrument and a computer
network.
1.
In Thermo Fisher
Connect Platform, click
for the Instrument Connect page.
2.
Select any of your registered instruments.
Chapter4
Calibrate and verify instrument performance
View the calibration status and set reminders
4
QuantStudio
3 and 5 RealTime PCR Systems Installation, Use, and Maintenance Guide
51
3.
In the Summary tab:
Click Calibrations to view the status of each calibration type.
(Optional) Click Calibration Reminders to set the calibration reminder time table and enter
the notification email address(es).
Note: More than one email address can receive the calibration reminders.
Note: The settings are automatically saved.
(Optional) In the Downloads section, click Maintenance Summary.pdf to download the
calibration status report.
4.
In the Calibrations History tab:
View the history of each calibration type.
(Optional) Click to download the calibration history report.
Perform ROI/uniformity, background, and dye calibrations
Workflow: Calibration
Note: The ROI and uniformity calibrations use the same calibration plate.
Perform an ROI calibration
Perform a uniformity calibration
You are automatically prompted to perform background calibration.
Perform a background calibration
Perform any time that ROI calibrations are current.
Perform system dye calibrations
Perform any time that ROI and background calibrations are current.
(Optional ) Perform custom dye calibrations (“Calibrate custom dyes” on page65)
Perform any time that ROI and background calibrations are current.
Chapter4Calibrate and verify instrument performance
Perform ROI/uniformity, background, and dye calibrations
4
52
QuantStudio
3 and 5 RealTime PCR Systems Installation, Use, and Maintenance Guide
Prepare a calibration plate
Materials required for calibration plate preparation
Plate(s) for the calibration that you are performing:
ROI plate (the same plate is used for ROI and uniformity calibrations)
Background calibration plate
Dye calibration plates
Note: We recommend calibrating with all Spectral Dye Calibrations Plates available for your
block configuration even if you are not using all the dyes in the plates.
Note: Do not discard the packaging for the calibration plates. Each calibration plate can be used
up to 3 times if the following conditions are met:
·
The plate is stored in its packing sleeve at –15 to –25°C.
·
The plate is used within 6 months after opening.
·
The plate is used before the plate expiry date
Centrifuge with plate adapter; buckets cleaned before use
Powder‐free gloves
Safety glasses
Thaw, vortex, and centrifuge a calibration plate
1.
Remove the calibration plate from the freezer, then thaw the plate in its packaging. Keep plates
protected from light until you perform the calibration.
Thaw each plate for 30 minutes.
Use each plate within 2hours of thawing.
IMPORTANT!
Do not remove the plate from its packaging until you are ready to use it. The
fluorescent dyes in the wells of calibration plates are photosensitive. Prolonged exposure to light
can diminish the fluorescence of the dyes.
2.
While wearing powder‐free gloves, remove the calibration plate from its packaging and retain the
packaging. Do not remove the optical film.
3.
Vortex the plate for 5seconds, then centrifuge at 750–1,000×g for 2 minutes.
4.
Confirm that the liquid in each well is at the bottom of the well and free of bubbles. If it is not,
centrifuge the plate again.
Chapter4Calibrate and verify instrument performance
Perform ROI/uniformity, background, and dye calibrations
4
QuantStudio
3 and 5 RealTime PCR Systems Installation, Use, and Maintenance Guide
53
IMPORTANT! Keep the bottom of the plate clean. Fluids and other contaminants on the bottom
of the plate can contaminate the sample block and cause an abnormally high background signal.
Perform calibrations
1.
In the instrument home screen:
Calibration Touch
ROI/Uniformity
[1]
Settings4Maintenance and Service4Calibrations4ROI and Uniformity
Background
[2]
Settings4Maintenance and Service4Calibrations4Custom4Background
Dye Settings4Maintenance and Service4Calibrations4Dye
[1]
Automatically followed by Background calibration.
[2]
Initiate via this route if performing Background calibration only.
2.
Follow the instructions on the screen to start the calibration.
Note: Dye calibration only: Select the Dye Plate to run, then touch Next.
3.
Load the plate.
a.
Touch to eject the instrument drawer.
b.
Load the plate onto the plate adaptor so that the following criteria are met.
Well A1 of the plate is positioned in the top-left corner of the plate adapter.
The barcode faces the front of the instrument.
IMPORTANT!
The instrument should be loaded and unloaded by trained operators who have
been warned of the moving parts hazard.
Note: (For 96-well 0.2-mL blocks only) Do not remove the black plate adapter before loading
a plate or tube strips. If used, tube strips can fit loosely in the adapter, but the heated cover
will apply the appropriate pressure to seat the tube strips securely in the adapter.
Note: The 384-well and 96-well Fast (0.1-mL) block configurations do not require a plate
adapter.
Chapter4
Calibrate and verify instrument performance
Perform ROI/uniformity, background, and dye calibrations
4
54
QuantStudio
3 and 5 RealTime PCR Systems Installation, Use, and Maintenance Guide
c.
Touch to close the instrument drawer.
4.
Touch Start.
5.
When the run is complete and the screen displays Calibration Complete, touch View Results to
check the calibration status.
Calibration status Action
Passed Touch Next to proceed to the next required calibration.
Failed See “Troubleshoot calibration failure” on page57.
Note:
You can view the calibration images only after the ROI/Uniformity and Background
calibrations pass.
6.
When the run ends, unload the plate.
a.
Touch to eject the instrument drawer.
b.
Remove the plate.
c.
Touch to close the instrument drawer.
CAUTION!
PHYSICAL INJURY HAZARD. During instrument operation, the plate temperature
can reach 100°C. Allow it to cool to room temperature before handling.
Note: If the instrument does not eject the plate, contact Support.
7.
Return the plate to its original packaging.
Note: Each calibration plate can be used up to 3 times if the following conditions are met:
·
The plate is stored in its packing sleeve at –15 to –25°C.
·
The plate is used within 6 months after opening.
·
The plate is used before the plate expiry date
Chapter4Calibrate and verify instrument performance
Perform ROI/uniformity, background, and dye calibrations
4
QuantStudio
3 and 5 RealTime PCR Systems Installation, Use, and Maintenance Guide
55
View calibration images and transfer results to a USB drive
The instrument performs the ROI, Uniformity, and Background calibrations in sequence. You can view
the calibration images after the Background calibration is complete.
If an Administrator has set data management as system-defined, you cannot select the location to save
the data. See “Configure instrument settings” on page27.
1.
In the Calibration Status screen, touch Details.
2.
In the Details screen, touch a calibration type to view its images and plots.
Calibration Example results indicating successful calibration
ROI
Note: Select the desired filter
combination from the Filter Set
dropdown list.
Green circles around all wells and bright well centers.
Uniformity Signals from each well following a uniform trend.
Background Few, if any, signals with abnormally high fluorescence.
Dye Signals from each well following a uniform trend, and each dye
peaks at the correct filter.
Chapter4Calibrate and verify instrument performance
Perform ROI/uniformity, background, and dye calibrations
4
56
QuantStudio
3 and 5 RealTime PCR Systems Installation, Use, and Maintenance Guide
3.
In the Calibration Status screen, touch Accept Results or Reject Results.
Accepting the results saves the calibration data to the instrument and overwrites existing data.
4.
(Optional) Touch Transfer EDS to transfer the calibration data to a USB drive.
Troubleshoot calibration failure
Observation Possible cause Recommended action
Calibration failed
The plate was improperly prepared. Ensure the following:
The correct plate was used for the calibration
performed.
The plate was properly thawed.
The plate was properly centrifuged.
The plate was properly sealed.
The plate is damaged or
contaminated.
Check for damage, improper plate seal, or
contamination.
Order a replacement plate. If the replacement
plate fails, contact Support.
High fluorescence
signal in individual wells
Signals that exceed the limit of
normal fluorescence may indicate
fluorescent contaminants on the
plate or the sample block.
See “Identify contamination” on page57.
Calibration failed but
plate is undamaged
The incorrect plate was used for
calibration performed.
Use the plate that matches the calibration
performed.
The plate was improperly prepared. Repeat the calibration with the plate properly
prepared.
If the calibration fails again, order a replacement
plate. If the replacement plate fails, contact
Support.
Identify contamination
Signals that exceed the limit of normal fluorescence may indicate fluorescent contaminants on the
calibration plate or the sample block. Common contaminants include ink residue from permanent pens,
powder from disposable gloves, and dust.
1.
View the calibration data and note the wells that failed the calibration.
2.
Remove the plate from the instrument, rotate the plate 180°, then perform the calibration again.
Chapter4Calibrate and verify instrument performance
Perform ROI/uniformity, background, and dye calibrations
4
QuantStudio
3 and 5 RealTime PCR Systems Installation, Use, and Maintenance Guide
57
3.
Determine the location of the failed wells again as in step1.
Position of failed wells Action
Identical The sample block is contaminated.
Decontaminate the sample block (see page77).
Reversed The plate is contaminated.
Discard the plate, then perform the calibration using a new calibration plate.
4.
If the calibration fails after you decontaminate the sample block and replace the plate, contact
Support.
Create a background plate (optional)
Whenever possible, use a background plate listed in Appendix C, “Parts and materials”. These plates
contain a buer that accurately simulates the reagents used for PCR, and, therefore, produces high-
quality calibration data.
If a background plate is not available, you can create one as described below.
Required materials:
MicroAmp
optical 96-well reaction plate
Optical adhesive cover or optical flat caps
Pipettor, 200‐µL (with pipette tips)
Powder‐free gloves
Safety glasses
Deionized water
IMPORTANT!
Wear powder-free gloves while creating the background plate.
1.
Remove a reaction plate from its box and place it on a clean, dry surface.
2.
Aliquot deionized water to each well of the reaction plate.
The recommended volume for a 96-well plate is 10-20µL.
The recommended volume for 384-well plate is 5µL.
3.
Seal the plate using an optical adhesive cover or optical flat caps.
4.
Use the plate for background calibration.
Chapter4Calibrate and verify instrument performance
Perform ROI/uniformity, background, and dye calibrations
4
58
QuantStudio
3 and 5 RealTime PCR Systems Installation, Use, and Maintenance Guide
Perform instrument verification using RNase P plates
Instruments are factory-calibrated, so calibration is not necessary for initial installation. However,
perform instrument verification:
After installation and before first use of the instrument.
After performing instrument calibrations. (recommended)
As needed to confirm instrument performance after calibrations, repair, preventative maintenance,
or change to the system.
The instrument requires valid ROI/uniformity, background, and dye calibrations to perform instrument
verification.
Instrument verification description
Purpose
Description Pass criteria
Confirms the
performance of
the instrument.
Quantifies the number of
copies of the human RNaseP
gene in samples with
known concentrations of the
corresponding DNA template.
The instrument passes performance specifications if the
following inequality is true and the instrument successfully
distinguishes between unknown populations AandB.
[(C
tA
) – 3(σ
CtA
)] > [(C
tB
) + 3(σ
CtB
)]
where:
C
tA
= Average C
t
of unknown populationA
σ
CtA
= Standard deviation of unknown populationA
C
tB
= Average C
t
of unknown populationB
σ
CtB
= Standard deviation of unknown populationB
RNaseP instrument verification plate
The RNase P plate contains the reagents necessary for the detection and quantitation of genomic
copies of the human RNaseP gene (a single-copy gene encoding the RNase moiety of the RNaseP
enzyme). Each well contains: PCR master mix, RNaseP primers, FAM
dye-labeled probe, and a
known concentration of human genomic DNA template.
Chapter4Calibrate and verify instrument performance
Perform instrument verification using RNase P plates
4
QuantStudio
3 and 5 RealTime PCR Systems Installation, Use, and Maintenance Guide
59
1
4
2
5
76
3
8
Figure1 96-well RNase P plate
1
Unknown A (5,000)
2
NTC (no template control)
3
STD 1,250 copies
4
STD 2,500 copies
5
STD 5,000 copies
6
STD 10,000 copies
7
STD 20,000 copies
8
Unknown B (10,000)
1
2
4
5
7
6
3
8
Figure2
384-well RNase P plate
1
NTC (no template control)
2
STD 1,250 copies
3
STD 2,500 copies
4
STD 5,000 copies
5
STD 10,000 copies
6
STD 10,000 copies
7
Unknown A (5,000)
8
Unknown B (10,000)
Chapter4Calibrate and verify instrument performance
Perform instrument verification using RNase P plates
4
60
QuantStudio
3 and 5 RealTime PCR Systems Installation, Use, and Maintenance Guide
Performance specifications pass criteria
After the run, the software calculates average copy number values and standard deviation values.
The instrument passes performance specifications if the following inequality is true and the instrument
successfully distinguishes between unknown populations AandB.
[(C
tA
) – 3(σ
CtA
)] > [(C
tB
) + 3(σ
CtB
)]
C
tA
= Average C
t
of unknown populationA
σ
CtA
= Standard deviation of unknown populationA
C
tB
= Average C
t
of unknown populationB
σ
CtB
= Standard deviation of unknown populationB
The software automatically adjusts the threshold and omits a defined number of wells from the
unknown populations to meet the performance specifications. To view any omitted wells, open the
EDS file for the verification in the desktop software.
Prepare an RNase P plate
Materials required for RNase P plate preparation
RNase P instrument verification plate
Centrifuge with plate adapter; buckets cleaned before use
Powder‐free gloves
Safety glasses
Thaw, vortex, and centrifuge an RNase P plate
IMPORTANT!
Expose the RNase P plate to room temperature for no more than 45 minutes, inclusive
of thawing and preparation time.
After thawing, the RNase P plate cannot be refrozen.
1.
Remove the RNase P plate from the freezer, then thaw the plate in its packaging.
Thaw the plate for approximately 5 minutes.
Use the plate within 30 minutes of thawing.
2.
Confirm that the bench, vortex, and centrifuge are clean. Before use, wipe the vortex and
centrifuge using a lintfree tissue.
3.
While wearing powder‐free gloves, remove the plate from its packaging.
4.
Vortex the plate for 5seconds, then centrifuge at 750–1,000×g for 2 minutes.
Chapter4Calibrate and verify instrument performance
Perform instrument verification using RNase P plates
4
QuantStudio
3 and 5 RealTime PCR Systems Installation, Use, and Maintenance Guide
61
5.
Confirm that the liquid in each well is at the bottom of the well and free of bubbles. If it is not,
centrifuge the plate again.
IMPORTANT! Keep the bottom of the plate clean. Fluids and other contaminants on the bottom
of the plate can contaminate the sample block and cause an abnormally high background signal.
Perform RNaseP verification
If an Administrator has set data management as system-defined, you cannot select the location to save
the data. See “Configure instrument settings” on page27.
1.
In the home screen, touch Settings4Maintenance and Service4RNase P Verification.
2.
Load the plate.
a.
Touch to eject the instrument drawer.
b.
Load the plate onto the plate adaptor so that the following criteria are met.
Well A1 of the plate is positioned in the top-left corner of the plate adapter.
The barcode faces the front of the instrument.
IMPORTANT!
The instrument should be loaded and unloaded by trained operators who have
been warned of the moving parts hazard.
Note: (For 96-well 0.2-mL blocks only) Do not remove the black plate adapter before loading
a plate or tube strips. If used, tube strips can fit loosely in the adapter, but the heated cover
will apply the appropriate pressure to seat the tube strips securely in the adapter.
Note: The 384-well and 96-well Fast (0.1-mL) block configurations do not require a plate
adapter.
Chapter4
Calibrate and verify instrument performance
Perform instrument verification using RNase P plates
4
62
QuantStudio
3 and 5 RealTime PCR Systems Installation, Use, and Maintenance Guide
c.
Touch to close the instrument drawer.
3.
Touch Start.
4.
When the run is complete and the screen displays Verification Complete, touch View Results to
confirm the status of the run.
Calibration status Action
Passed Instrument is ready for use.
Failed See “Troubleshoot verification failure” on page63.
5.
In the RNase P Verification Status screen, touch:
Accept Results to save the results to the instrument
Reject Results to delete the RNaseP verification results
Export Results to export the calibration results to a USB drive
6.
When the run ends, unload the plate.
a.
Touch to eject the instrument drawer.
b.
Remove the plate.
c.
Touch to close the instrument drawer.
CAUTION!
PHYSICAL INJURY HAZARD. During instrument operation, the plate temperature
can reach 100°C. Allow it to cool to room temperature before handling.
Note:
If the instrument does not eject the plate, contact Support.
Troubleshoot verification failure
Observation Possible cause Recommended action
Verification failed
The plate was improperly
prepared.
Ensure the following:
The correct plate was used for the verification
performed.
The plate was properly thawed.
The plate was properly centrifuged.
The plate was properly sealed.
Open the data file for the verification in the desktop
software or the Thermo Fisher
Connect Platform to view
the flags and troubleshooting details for failed wells.
The plate is damaged or
contaminated.
Check for damage, improper plate seal, or contamination.
Order a replacement plate. If the replacement plate fails,
contact Support.
Chapter4Calibrate and verify instrument performance
Perform instrument verification using RNase P plates
4
QuantStudio
3 and 5 RealTime PCR Systems Installation, Use, and Maintenance Guide
63
Observation Possible cause Recommended action
High fluorescence
signal
The reaction volume is not
correct.
Ensure that reaction volumes in the plate are correct and
match the volume that is entered in the Method tab.
Signals that exceed the
limit of normal fluorescence
may indicate fluorescent
contaminants on the plate or
the sample block.
Examine the bottom of the reaction plate. If there is
contamination, prepare a new plate and run the experiment
again.
Identify the location of contamination on the plate or sample
block.
1.
Obtain a background plate.
2.
Follow the procedures that are described in “Identify
contamination” on page57.
Verification failed
but plate is
undamaged
The incorrect plate was used
for verification.
Use the correct RNase P plate for verification.
The plate was improperly
prepared.
Repeat the verification with a new properly prepared plate.
Note: The verification procedure is an experiment run, so
each RNase P plate can only be used once.
Open the data file for the verification in the desktop
software or the Thermo Fisher
Connect Platform to view
the flags and troubleshooting details for failed wells.
If the verification fails again, order a replacement plate. If
the replacement plate fails, contact Support.
Chapter4Calibrate and verify instrument performance
Perform instrument verification using RNase P plates
4
64
QuantStudio
3 and 5 RealTime PCR Systems Installation, Use, and Maintenance Guide
Calibrate custom dyes
Custom dyes overview
The instrument can run assays designed with custom dyes. Custom dyes include:
Dyes that are not manufactured by Thermo Fisher Scientific.
Dyes or formulations of dyes that are not system dyes for the instrument.
For custom dye requirements, see “Custom dyes” on page14.
Workflow: Custom dye calibration
For each custom dye, determine the optimal dye concentration. Use this concentration to prepare all
subsequent dye calibration plates.
Use a dilution series to determine an optimal custom dye concentration
Prepare a custom dye dilution plate
Run the dilution plate as an experiment
Determine the optimal dye concentration
Calibrate the custom dye using the optimal concentration
Create a custom dye calibration plate
Add a custom dye to the instrument
Perform a custom dye calibration
Use a dilution series to determine an optimal custom dye concentration
Custom dye dilution guidelines
Prepare a dilution series for each custom dye.
Target several dye concentrations within a range of 100–2,000 nM.
Choose a 2 or 3fold dierence in dilution points.
Dispense 10–20µL per well for 96well plates or 5µL per well for 384well plates.
Dilute the dye in buer compatible with your master mix.
(Intercalating dyes only) Add the appropriate amount of amplified PCR product to generate
fluorescence.
Chapter4
Calibrate and verify instrument performance
Calibrate custom dyes
4
QuantStudio
3 and 5 RealTime PCR Systems Installation, Use, and Maintenance Guide
65
Prepare a custom dye dilution plate
IMPORTANT! Wear powder-free gloves throughout the procedure.
1.
Prepare a 2- or 3-fold dilution series of the custom dye.
2.
Dispense aliquots of each dilution into the center of a reaction plate, then seal the plate.
A full plate is not needed. See the following figures for suggested replicates.
1
2
3
4
5
1
100nM
2
200nM
3
400nM
4
800nM
5
1,600nM
1
7
6
5
4
3
2
8 9 10
1
100nM
2
200nM
3
400nM
4
800nM
5
1,600nM
6
100nM
7
200nM
8
400nM
9
800nM
10
1,600nM
3.
Vortex the plate for 5seconds, then centrifuge at 750–1,000×g for 2 minutes.
Chapter4Calibrate and verify instrument performance
Calibrate custom dyes
4
66
QuantStudio
3 and 5 RealTime PCR Systems Installation, Use, and Maintenance Guide
4.
Confirm that the liquid in each well is at the bottom of the well and free of bubbles. If it is not,
centrifuge the plate again.
IMPORTANT! Keep the bottom of the plate clean. Fluids and other contaminants on the bottom
of the plate can contaminate the sample block and cause an abnormally high background signal.
Run the dilution plate as an experiment
If an Administrator has set data management as system-defined, you cannot select the location to save
the data. See “Configure instrument settings” on page27.
1.
Load the plate into the instrument.
CAUTION!
The instrument should be used by trained operators who have been warned of the
moving parts hazard.
2.
Set up a genotyping experiment on the instrument.
Load an experiment created in the desktop or cloud software.
a.
In the home screen, touch Load Experiment.
b.
Select the experiment location, then the experiment file.
Create a new experiment on the instrument touchscreen.
a.
In the home screen, touch Open Template4Genotyping4Genotyping Post.
b.
(Optional) In the Properties tab, edit the experiment properties.
c.
In the Method tab, set the hold temperature to 60°C with a 2 minute hold and enter the
appropriate reaction volume.
d.
In the Plate tab, enter the dilution series information for the appropriate wells.
3.
Touch Start Run.
4.
When the run is complete, download the results for analysis.
5.
Unload the plate from the instrument.
CAUTION!
PHYSICAL INJURY HAZARD. During instrument operation, the plate temperature
can reach 100°C. Allow it to cool to room temperature before handling.
Note: If the instrument does not eject the plate, contact Support.
Chapter4
Calibrate and verify instrument performance
Calibrate custom dyes
4
QuantStudio
3 and 5 RealTime PCR Systems Installation, Use, and Maintenance Guide
67
Determine the optimal dye concentration
Review the dye signal data and select the dilution to use for dye calibration.
1.
In the Results tab of the desktop or cloud software, select Raw Data Plot.
This plot displays the raw fluorescence signal of each optical filter, for individual wells.
Note: The Raw Data Plot cannot be viewed on the instrument touchscreen.
2.
For each replicate population of dilutions, select the wells in the Plate Layout to view in the plot.
3.
Examine the raw data to identify the wells yielding signals according to the ranges shown in the
following table.
Plate type
Acceptable signal range
[1]
96-well 800,000 to 3,200,000
384-well 400,000 to 2,000,000
[1]
Signal range for the optical filter where the dye is brightest.
4.
(Optional) Export the raw data, then calculate the average fluorescence value for each
concentration.
5.
Select the lowest (optimal) dye concentration that falls within the acceptable signal range.
Calibrate the custom dye
Create a custom dye calibration plate
IMPORTANT!
Wear powder-free gloves while creating the dye plate.
Create a full plate of the custom dye diluted to the optimal concentration:
1.
Dilute the custom dye to the optimal concentration in buer.
Prepare an adequate volume, using a volume range of 10–20µL/well for a 96well plate or
5 µL/well for a 384-well plate.
2.
Pipet the appropriate volume of the diluted custom dye to the plate wells.
3.
Seal the plate.
4.
Vortex the plate for 5seconds, then centrifuge at 750–1,000×g for 2 minutes.
Chapter4Calibrate and verify instrument performance
Calibrate custom dyes
4
68
QuantStudio
3 and 5 RealTime PCR Systems Installation, Use, and Maintenance Guide
5.
Confirm that the liquid in each well is at the bottom of the well and free of bubbles. If it is not,
centrifuge the plate again.
IMPORTANT! Keep the bottom of the plate clean. Fluids and other contaminants on the bottom
of the plate can contaminate the sample block and cause an abnormally high background signal.
Add a custom dye to the instrument
1.
In the instrument home screen, touch Settings4Maintenance and
Service4Calibrations4Custom4Custom Dye.
2.
Touch Add Custom Dye.
3.
Enter the dye information:
Field/option
Action
Custom Dye
Name
Enter a name for the custom dye.
IMPORTANT!
·
Do not use a system dye name for a custom dye name.
·
Dye names are spacing sensitive and cannot contain special characters.
Type Select:
Reporter—The dye works in conjunction with a quencher dye to report an
increase of PCR product.
Quencher—The dye suppresses the fluorescence of a reporter dye until
amplification of PCR product.
Both—The dye can be used as a reporter or quencher dye.
4.
Touch Save.
IMPORTANT!
You must also add the custom dye to the desktop software dye library before creating,
running, or analyzing experiments that use the custom dye. For detailed information, see the documents
for the desktop software.
Chapter4Calibrate and verify instrument performance
Calibrate custom dyes
4
QuantStudio
3 and 5 RealTime PCR Systems Installation, Use, and Maintenance Guide
69
Perform a custom dye calibration
If an Administrator has set data management as system-defined, you cannot select the location to save
the data. See “Configure instrument settings” on page27.
1.
Load the plate into the instrument.
CAUTION! The instrument should be used by trained operators who have been warned of the
moving parts hazard.
2.
In the instrument home screen, touch Settings4Maintenance and
Service4Calibrations4Custom4Custom Dye.
3.
Touch the custom dye to calibrate.
4.
Review the custom dye information, (optional) make changes, then touch Update.
5.
Enter the calibration temperature.
6.
(Optional) Touch Reagents, then enter reagent information.
7.
Touch Start.
8.
When the run is complete and the screen displays Calibration Complete, touch View
Results4Details.
9.
Review the plot. Passing calibration results show uniform signals with peaks that are aligned with
the dye wavelength.
Peak
filter
Filter wavelength (nm)
[1]
Excitation Emission
x1-m1 470 ± 15nm 520 ± 15nm
x2-m2 520 ± 10nm 558 ± 12nm
x3-m3 550 ± 10nm 587 ± 10nm
x4-m4 580 ± 10nm 623 ± 14nm
Chapter4Calibrate and verify instrument performance
Calibrate custom dyes
4
70
QuantStudio
3 and 5 RealTime PCR Systems Installation, Use, and Maintenance Guide
(continued)
Peak filter
Filter wavelength (nm)
[1]
Excitation Emission
x5-m5 640 ± 10nm 682 ± 14nm
x6-m6 662 ± 10nm 711 ± 12nm
[1]
The central wavelengths are the optimized wavelengths.
Note: Example dye calibration plot.
The peaks for your dye may align with a dierent filter set.
10.
Select an action depending on whether the custom dye calibration passed or failed.
Calibration
status
Action
Passed
Touch Accept Results or Reject Results.
Note: Accepting the results saves the calibration data to the instrument and
overwrites existing data.
(Optional) Touch Transfer EDS to transfer the calibration data to a USB.
Failed
Perform the calibration again using a new custom dye plate.
See “Troubleshoot calibration failure” on page57.
11.
Unload the plate from the instrument.
CAUTION!
PHYSICAL INJURY HAZARD. During instrument operation, the plate temperature
can reach 100°C. Allow it to cool to room temperature before handling.
Note: If the instrument does not eject the plate, contact Support.
Chapter4
Calibrate and verify instrument performance
Calibrate custom dyes
4
QuantStudio
3 and 5 RealTime PCR Systems Installation, Use, and Maintenance Guide
71
Calibrate for a custom melt curve experiment
A custom melt calibration calibrates a custom dye and a melt calibration at the same time.
If an Administrator has set data management as system-defined, you cannot select the location to save
the data. See “Configure instrument settings” on page27.
Before performing a calibration for a custom melt curve experiment, ensure that all calibrations are
current (touch Settings4Maintenance and Service4Calibrations4History and Reminders).
Add a custom dye to the software (see “Add a custom dye to the instrument” on page69).
1.
In the home screen, touch Settings4Maintenance and
Service4Calibrations4Custom4Custom Melt.
2.
Touch PCR + Melt or Melt only as appropriate for the kit you are using.
3.
Load the plate into the instrument.
4.
Select or add a dye, then select a filter set appropriate for your dye's wavelength (see filter-
wavelength table below).
Note: See your reagent kit documentation for dye name and wavelength information.
Peak
filter
Filter wavelength (nm)
[1]
Excitation Emission
x1-m1 470 ± 15nm 520 ± 15nm
x2-m2 520 ± 10nm 558 ± 12nm
x3-m3 550 ± 10nm 587 ± 10nm
x4-m4 580 ± 10nm 623 ± 14nm
x5-m5 640 ± 10nm 682 ± 14nm
x6-m6 662 ± 10nm 711 ± 12nm
[1]
The central wavelengths are the optimized wavelengths.
IMPORTANT!
If the selected filter set does not match your reagent kit documentation, then the
incorrect wavelength may be collected during a run.
5.
(Optional) Touch Reagents, then enter reagent information.
6.
Touch Start.
7.
When the run is complete and the screen displays Calibration Complete, touch View
Results4Details.
Chapter4
Calibrate and verify instrument performance
Calibrate for a custom melt curve experiment
4
72
QuantStudio
3 and 5 RealTime PCR Systems Installation, Use, and Maintenance Guide
8.
Review the plot. Passing calibration results show uniform signals with peaks that are aligned with
the dye wavelength.
Note: Example dye calibration plot.
The peaks for your dye may align with a dierent filter set.
9.
Select an action depending on whether the custom dye calibration passed or failed.
Calibration
status
Action
Passed
Touch Accept Results or Reject Results.
Note: Accepting the results saves the calibration data to the instrument and
overwrites existing data.
(Optional) Touch Transfer EDS to transfer the calibration data to a USB.
Failed
Perform the calibration again using a new custom dye plate.
See “Troubleshoot calibration failure” on page57.
10.
Unload the plate from the instrument.
Note: You must also add the custom dye to the desktop software dye library before creating, running,
or analyzing experiments that use the custom dye. For detailed information, see the documentation for
the desktop software.
Note: To perform a custom melt experiment, you can either create a Standard Curve or a Custom
experiment with melt, then specify the data points per degree in the method.
Chapter4Calibrate and verify instrument performance
Calibrate for a custom melt curve experiment
4
QuantStudio
3 and 5 RealTime PCR Systems Installation, Use, and Maintenance Guide
73
Maintain the instrument
Back up or restore theinstrument ...................................................... 74
Perform a self-verification test .......................................................... 75
View instrumentlogs .................................................................. 75
Configure the networkdrive ............................................................ 75
Decontaminate the sampleblock ....................................................... 77
Replace the instrumentfuses .......................................................... 80
Power on or o, store, andmove ....................................................... 81
IMPORTANT!
This chapter contains user maintenance procedures for the instrument. Procedures
other than those described in this document must be performed by a qualified Thermo Fisher Scientific
service representative.
Back up or restore the instrument
In the home screen, touch Settings4Maintenance and Service4Backup/Restore.
To
Action
Back up
1.
Touch Backup Instrument.
2.
Select a location to store the backup settings for your instrument.
3.
If backing up to a USB drive, insert a USB drive into the front-panel USB port.
4.
Enter a backup file name, then touch Done.
5.
Select the elements to back up, or leave them all selected.
6.
Touch Backup.
Restore (Administrator
only)
1.
Touch Restore a Backup.
2.
Select the storage location of the backup settings.
3.
If restoring from a USB drive, insert a USB drive into the front-panel USB port.
4.
Select the backup file, then touch Restore.
5
74
QuantStudio
3 and 5 RealTime PCR Systems Installation, Use, and Maintenance Guide
Perform a self-verification test
If an Administrator has set data management as system-defined, you cannot select the location to save
the data. See “Configure instrument settings” on page27.
1.
In the home screen, touch Settings4Maintenance and Service4Self Verification Test.
IMPORTANT! Ensure the front of the instrument is not obstructed. The instrument drawer will be
ejected during the test. The test will take approximately ten minutes.
2.
Touch Start Test.
View instrument logs
If an Administrator has set data management as system-defined, you cannot select the location to save
the data. See “Configure instrument settings” on page27.
1.
In the home screen, touch Settings4Maintenance and Service4Log.
2.
Touch Instrument Run Log to view the log records.
3.
(Optional) Touch a category on the left to filter the log table by event type.
4.
(Optional) Touch an individual log record to view details.
5.
(Optional) Export log records.
Touch Export4 USB.
Touch Export4 Thermo Fisher
Connect Platform, then touch Export.
Touch Export4 Network.
Configure the network drive
When importing or exporting files from and to a network drive, you are prompted to specify the network
drive location and information.
If the instrument is already connected to a network drive, the folder structure of the network drive is
displayed.
The network drive can be identified by the IP address of the computer. It can be identified by the
domain name system (DNS) name for v1.5.2 or later.
For v1.6 or later, the location of the network drive can be set in the data management settings. If the
network drive is defined in the data management settings, it cannot be changed at the time that files are
exported. For more information, see “Configure instrument settings” on page27.
Set up the network drive as a shared folder (see “Create a shared network folder” on page76).
If the network drive is identified by the IP address of the computer, determine the IP address of the
computer (see “Determine the IP address of the computer” on page77).
Chapter5
Maintain the instrument
Perform a self-verification test
5
QuantStudio
3 and 5 RealTime PCR Systems Installation, Use, and Maintenance Guide
75
If the network drive is identified by the DNS name, determine the DNS name (see “Determine the DNS
name” on page77).
1.
In the home screen, touch Settings4Instrument Settings4Network Drive.
2.
Select the drive location.
The drive location is required.
For IP address, the drive location is <IP address>/<folder name>.
For DNS name, the drive location is <DNS>/<folder name>.
3.
Enter the following information, then touch Connect.
Domain name
User name
Password
The domain name, username, and password can be required, depending on the network
requirements.
The username and password are for the network. They are not the username and password for the
instrument.
Create a shared network folder
Files can be transferred from the instrument to a folder on the network.
The folder on the network must be set up as a shared folder.
1.
To designate a folder as shared, right-click on the folder, then click Properties.
2.
In the Properties dialog box, click the Sharing tab.
3.
Click Advanced Sharing.
4.
In the Advanced Sharing dialog box, select the Share this folder checkbox, then click
Permissions.
5.
Ensure that Everyone is selected in the Group or user name field.
6.
Ensure that the checkboxes for Allow are selected for the permissions.
The permissions are Full Control, Change, and Read.
7.
In the Permissions dialog box, click OK.
8.
In the Advanced Sharing dialog box, click Apply4OK.
9.
In the Properties dialog box, click Close.
Chapter5
Maintain the instrument
Configure the network drive
5
76
QuantStudio
3 and 5 RealTime PCR Systems Installation, Use, and Maintenance Guide
Determine the IP address of the computer
1.
In Windows
search bar, enter cmd to open the Command Prompt.
2.
In the Command Prompt, enter ipconfig/all, then press Enter.
3.
Locate the IPv4 Address.
The IP address is listed on the IPv4 Address line.
Determine the DNS name
1.
In the Windows
search bar, enter
cmd
to open the Command Prompt.
2.
In the Command Prompt, enter ipconfig/all, then press Enter.
3.
Locate the DNS names.
The text indicating the DNS names might be dierent, depending on the configuration. The text
DNS is displayed and there might be more than one field for the DNS names. Contact your IT
administrator for assistance to determine the correct DNS name.
Decontaminate the sample block
Perform this procedure to eliminate fluorescent contaminants from the instrument sample block.
Contamination is generally evident in failed background calibrations where one or more wells
consistently exhibit abnormally high signals.
CAUTION!
PHYSICAL INJURY HAZARD. Do not remove the instrument cover. There are no
components inside the instrument that you can safely service yourself. If you suspect a problem,
contact Support.
CAUTION! PHYSICAL INJURY HAZARD. During instrument operation, the sample block
temperature can reach 100°C. Allow it to cool to room temperature before handling.
CAUTION! Before using a cleaning or decontamination method other than those recommended by
Thermo Fisher Scientific, confirm with Thermo Fisher Scientific that the proposed method will not
damage the equipment.
Materials required
Safety glasses
Powder-free gloves
Tissue, lint-free
Cotton or nylon swabs and lint-free cloths
Pipette (100-µL) with pipette tips
Chapter5
Maintain the instrument
Decontaminate the sample block
5
QuantStudio
3 and 5 RealTime PCR Systems Installation, Use, and Maintenance Guide
77
Deionized water
Ethanol, 95% solution
Na-hypochlorite, (0.1% v/v) solution
Clean the sample block
CAUTION! PHYSICAL INJURY HAZARD. During instrument operation, the sample block
temperature can reach 100°C. Allow it to cool to room temperature before handling.
IMPORTANT! Wear powder-free gloves when you perform this procedure.
IMPORTANT! Always use deionized water to rinse wells after cleaning with bleach or ethanol solution.
1.
Identify the contaminated wells of the sample block (see “Identify contamination” on page57).
2.
Prepare the instrument and access the sample block:
a.
Power o and unplug the instrument, then allow it to cool for 15 minutes.
b.
Touch to eject the instrument drawer.
3.
Rinse the contaminated wells with deionized water (see “Detailed procedures for cleaning the
sample block” on page79).
4.
Close the drawer and test the sample block for contamination:
a.
Push the instrument drawer back in to the instrument.
b.
Plug in, then power on the instrument.
c.
Perform a background calibration to confirm that you have eliminated the contamination.
5.
If the contamination remains, clean the contaminated wells using a 95% ethanol solution:
a.
Repeat step 2 – step 3.
b.
Clean the contaminated wells using a 95% ethanol solution (see “Detailed procedures for
cleaning the sample block” on page79).
Chapter5
Maintain the instrument
Decontaminate the sample block
5
78
QuantStudio
3 and 5 RealTime PCR Systems Installation, Use, and Maintenance Guide
c.
Repeat step 3 – step 4 to rinse the sample block with deionized water and to confirm that you
have eliminated the contamination.
IMPORTANT! Always use deionized water to rinse wells after cleaning with bleach or ethanol
solution.
6.
If the contamination still remains, clean the contaminated wells using a Na-hypochlorite (0.1% v/v)
solution:
a.
Repeat step 2 – step 3.
b.
Clean the contaminated wells using a Na-hypochlorite (0.1% v/v) solution (see “Detailed
procedures for cleaning the sample block” on page79).
c.
Repeat step 3 – step 4 to rinse the sample block with deionized water and to confirm that you
have eliminated the contamination.
IMPORTANT!
Always use deionized water to rinse wells after cleaning with Na-hypochlorite
or ethanol solution.
7.
If the contamination continues to remain, contact Support.
Detailed procedures for cleaning the sample block
IMPORTANT! Use these cleaning procedures only in conjunction with the complete decontamination
procedure (see “Clean the sample block” on page78).
Rinse the sample block with deionized water.
a.
Pipet a small volume of deionized water into each contaminated well.
b.
In each well, pipet the water up and down several times to rinse the well.
c.
Pipet the water to a waste beaker.
d.
Use a cotton swab to scrub inside of each contaminated well.
e.
Use a lint-free cloth to absorb the excess deionized water.
Clean the sample block with 95% ethanol.
a.
Pipet a small volume of 95% ethanol solution into each contaminated well.
b.
In each well, pipet the solution up and down several times to rinse the well.
c.
Pipet the ethanol solution to a waste beaker.
IMPORTANT!
Always use deionized water to rinse wells after cleaning with bleach or ethanol
solution.
Clean the sample block with Na-hypochlorite (0.1% v/v) solution.
a.
Pipet a small volume of Na-hypochlorite (0.1% v/v) solution into each contaminated well.
Chapter5
Maintain the instrument
Decontaminate the sample block
5
QuantStudio
3 and 5 RealTime PCR Systems Installation, Use, and Maintenance Guide
79
b.
In each well, pipet the solution up and down several times to rinse the well.
c.
Pipet the bleach solution to a waste beaker.
IMPORTANT! Always use deionized water to rinse wells after cleaning with bleach or ethanol
solution.
Replace the instrument fuses
CAUTION! FIRE HAZARD. For continued protection against the risk of fire, replace fuses only with
listed and certified fuses of the same type and rating as those currently in the instrument.
Materials required
Fuses (2) – 10A, Time-Lag T, 250VAC, 5×20mm
Safety glasses
Powder-free gloves
Screwdriver, flathead
Replace the fuses
1.
Power o and unplug the instrument, then allow it to cool for 15 minutes.
2.
Using a flat-head screwdriver, unscrew and remove the fuse holder.
3.
Remove each fuse from its fuse holder and inspect it for damage. Carbon typically coats the inside
of failed fuses.
Good
Failed
Chapter5Maintain the instrument
Replace the instrument fuses
5
80
QuantStudio
3 and 5 RealTime PCR Systems Installation, Use, and Maintenance Guide
4.
Replace each failed fuse.
Note: The voltage and amperage ratings are on the fuse holder.
5.
Install the fuse holder back into the instrument.
6.
Plug in, then power on the instrument.
The installation is successful if the instrument powers on.
Note: Fuse failure can result from fluctuations in the supplied power to the system. To prevent further
failures, consider installing an electrical protective device, such as a UPS or a surge protector. If issues
with the fuse persist, contact Support.
Power on or o, store, and move
Enable sleep mode
1.
In the home screen, touch Settings4Instrument Settings4Sleep Mode.
2.
Slide the control On to enable sleep mode.
3.
Touch Edit Time, then enter the time length of inactivity before the instrument enters sleep mode.
4.
Touch Enter, then touch OK.
Power on the instrument
1.
Touch anywhere on the touchscreen to determine if the instrument is in sleep mode. If the home
screen is displayed, the instrument is already powered on.
2.
If the home screen does not display, power on the
instrument by pressing the switch on the rear panel.
If left unattended, the instrument automatically enters sleep
mode (enabled by default) to conserve power.
Note: To customize the sleep mode setting, touch
Settings4Instrument Settings4Sleep Mode.
Chapter5Maintain the instrument
Power on or o, store, and move
5
QuantStudio
3 and 5 RealTime PCR Systems Installation, Use, and Maintenance Guide
81
Power o the instrument
The instrument operates in low-power mode when not in use. However, the instrument can be powered
o completely so that the components use no power.
Note: To power o the instrument for >1 week, see “Prepare the instrument to store, move, or ship” on
page82.
1.
Power o the instrument using the power switch on the back of the instrument.
2.
Power o the computer.
Prepare the instrument to store, move, or ship
In the home screen:
1.
Touch Settings4Maintenance and Service4Ship Prep Mode4Next.
2.
Touch to eject the instrument drawer.
3.
Load the packing plate or an empty plate, then touch to close the drawer.
4.
Touch Lock Block.
5.
Power o the instrument using the power switch on the back of the instrument.
The instrument is now ready to store, move, or ship.
Move the instrument
CAUTION! PHYSICAL INJURY HAZARD. Do not attempt to lift the instrument or any other heavy
objects unless you have received related training. Incorrect lifting can cause painful and sometimes
permanent back injury. Use proper lifting techniques when lifting or moving the instrument. At least
two people are required to lift it.
IMPORTANT! Moving your instrument can create subtle changes in the alignment of the instrument
optics. Recalibrate the instrument if necessary.
Ensure that the surface on which you place the instrument can support at least 35kg (77lbs).
Ensure that the path to transport the instrument is clear of obstructions.
At least two people are needed to lift and carry the instrument.
Keep your spine in a good neutral position.
Bend at the knees and lift with your legs.
Do not lift an object and twist your torso at the same time.
Coordinate your intentions with your assistant before lifting and carrying.
Chapter5
Maintain the instrument
Power on or o, store, and move
5
82
QuantStudio
3 and 5 RealTime PCR Systems Installation, Use, and Maintenance Guide
IMPORTANT! After moving the instrument, perform an RNase P instrument verification run. If the run
fails, perform ROI/ uniformity, background, and dye calibrations.
Return the instrument for service
The service process requires 2 to 3 weeks.
Before returning the instrument for service, perform the following tasks.
1.
Back up the instrument (see “Back up or restore the instrument” on page74).
2.
In the home screen, touch Settings4Instrument Settings4Reset Factory defaults.
3.
Set the instrument to Ship Prep Mode (see “Prepare the instrument to store, move, or ship” on
page82).
To return the instrument for service, perform the following tasks.
1.
Contact your local customer care center or technical support group to obtain a copy of the
Certificate of Instrument Decontamination, a service notification, a service call number, and
packaging materials (if required).
2.
Follow the instructions in the form to decontaminate the instrument.
IMPORTANT!
The instrument must be decontaminated before packing it for shipping.
3.
Complete and sign a copy of the Certificate of Instrument Decontamination.
4.
Fax the Certificate of Instrument Decontamination to the customer care center.
5.
Pack the instrument in the provided packaging and follow the instructions in the table below.
Prepare
Include Exclude
1.
Transfer any data files from the
instrument.
2.
Load an empty plate in the
sample block.
3.
Use the touchscreen to place
the instrument in ship mode.
Note: The empty plate and
ship mode protects the internal
components of the instrument during
transport.
Instrument
Completed and signed
Certificate of Instrument
Decontamination
Note: The instrument will not be
accepted for service without a
hard copy of the Certificate of
Instrument Decontamination.
Any accessories, including:
Power cord
Ethernet cable
USB drive
Wireless adapter
Note: If included with
the instrument, these items
will be disposed of during
service and not returned.
6.
Attach the postage provided with the Certificate of Instrument Decontamination to the box, then
ship the instrument to the designated facility.
Chapter5
Maintain the instrument
Power on or o, store, and move
5
QuantStudio
3 and 5 RealTime PCR Systems Installation, Use, and Maintenance Guide
83
Configure the instrument and
manage instrument profiles
Initial startup ........................................................................ 84
Manageprofiles ...................................................................... 85
Link an instrument profile to Thermo Fisher
Connect .................................... 86
Enable SAE mode (Administratoronly) .................................................. 92
Require instrument profile sign-in (Administratoronly) ..................................... 93
Enable remote instrument monitoring (Administratoronly) ................................. 93
Update instrument software (Administratoronly) .......................................... 94
Restore factorydefaults ............................................................... 94
Configure thenetwork ................................................................ 94
Select a region for the Thermo Fisher
Connect Platform (Administratoronly) ............... 96
Manage the instrument name (Administratoronly) ........................................ 96
Set the date and time (Administratoronly) ............................................... 96
Manage the signout timer (Administrator only) ........................................... 97
Initial startup
Perform these tasks during initial instrument startup or after you restore factory defaults (see “Restore
factory defaults” on page94).
1.
(Optional) Select one of the following options.
Configure the instrument for either wired connection or wireless network connection (see
“Configure the network” on page94).
Touch Neither/Decide Later.
2.
(Optional) Enter an instrument name (see “Manage the instrument name (Administrator only)” on
page96).
3.
Set the date and time (see “Set the date and time (Administrator only)” on page96).
4.
Select one of the following options.
Select a region for the Thermo Fisher
Connect Platform.
Touch Next to continue.
6
84
QuantStudio
3 and 5 RealTime PCR Systems Installation, Use, and Maintenance Guide
5.
Create an administrator profile (see “Create an administrator instrument profile during initial start-
up” on page85).
6.
Select one of the following options.
Link the instrument profile to the Thermo Fisher
Connect Platform (see “Link an instrument
profile to Thermo Fisher
Connect” on page86).
Touch Skip.
Manage profiles
To
See
Create a profile
“Create an administrator instrument profile during initial start-up” on page85
“Create a new instrument profile” on page26
Configure a profile
“Link an instrument profile to Thermo Fisher
Connect” on page86
“Edit an instrument profile” on page86
View or manage all profiles
“Manage all instrument profiles (Administrator only)” on page85
Create an administrator instrument profile during initial start-up
During initial start-up, the user is automatically prompted to create an administrator instrument profile.
See “Initial startup” on page84.
The first instrument profile that is created during installation is given administrator privileges.
Administrators can grant administrative privileges to other users. See “Manage all instrument
profiles (Administrator only)” on page85.
1.
Touch Name, enter a username, then touch Done.
2.
Touch PIN, enter a four-digit numerical password, then touch Enter.
Note: Touch the Show PIN checkbox to switch PIN display on or o.
3.
Touch Confirm PIN, then repeat step2.
4.
Touch Create profile.
Manage all instrument profiles (Administrator only)
1.
In the home screen, access the All Profiles tab.
Touch My Profile4All Profiles.
Touch Settings4Manage Users4Manage Profiles4All Profiles.
A list of users, the date the profile was created, and the user type displays.
Chapter6
Configure the instrument and manage instrument profiles
Manage profiles
6
QuantStudio
3 and 5 RealTime PCR Systems Installation, Use, and Maintenance Guide
85
2.
Select the instrument profile to edit.
3.
Edit the profile.
To delete the profile, touch Delete profile4Delete.
To reset the PIN, touch Reset PIN4Reset.
The user will be directed to enter a new PIN at the next sign in.
To enable or disable administrative privileges, slide the control to Administrator or Standard,
respectively.
4.
Touch Done.
Edit an instrument profile
1.
In the home screen, touch MyProfile.
Note: Administrators can also navigate to this screen by touching Settings4Manage
Users4Manage Profiles.
2.
Touch Edit.
3.
Select the fields to edit, then make changes.
4.
Touch Done.
Link an instrument profile to Thermo Fisher
Connect
Linking your instrument profile to Thermo Fisher
Connect cloud-based platform allows you to access
the following functions:
View an instrument status from the cloud.
Download templates from your cloud storage to the instrument.
Transfer run data from the instrument to your cloud storage.
To link an instrument profile to Thermo Fisher
Connect, see the following sections:
“Recommended order to set up profiles” on page87
“Overview of local instrument profiles and Thermo Fisher
Connect profiles” on page87
“Thermo Fisher
Connect Platform instrument profile roles and functions” on page88
“Link the instrument to your Thermo Fisher
Connect Platform account” on page88
“Link a local profile to a Thermo Fisher
Connect profile” on page89
“Unlink a Thermo Fisher
Connect account” on page91
“If you link when you are signed in to the instrument” on page91
“Change the PIN for a Thermo Fisher
Connect profile” on page92
Chapter6
Configure the instrument and manage instrument profiles
Link an instrument profile to Thermo Fisher
Connect
6
86
QuantStudio
3 and 5 RealTime PCR Systems Installation, Use, and Maintenance Guide
Recommended order to set up profiles
Task
Description
Create a profile for the
instrument administrator.
The first profile on the instrument becomes the instrument administrator.
See “Create an administrator instrument profile during initial start-up” on page85.
Select a Thermo Fisher
Connect administrator.
A Thermo Fisher
Connect administrator is distinct from an instrument administrator.
The first Thermo Fisher
Connect profile to be linked to the instrument is automatically
assigned the role of a Thermo Fisher
Connect administrator. It is recommended that
the first Connect profile to be linked is the one that should be a Thermo Fisher
Connect administrator.
The roles can be updated at a later time on the Thermo Fisher
Connect platform.
Each person creates a
profile.
Create a local instrument profile. Thermo Fisher
Connect -enabled features are not
available.
(Optional) Link a local instrument profile to a Thermo Fisher
Connect profile.
Sign in with a Thermo Fisher
Connect profile. Thermo Fisher
Connect-enabled
features are available.
Assign additional
instrument administrators.
Any of the profiles can be assigned the role of instrument administrator. The local
profile must be created first or the Thermo Fisher
Connect profile must be linked first.
Overview of local instrument profiles and Thermo Fisher
Connect profiles
Local instrument profile
Experiment template files and data files are stored on the instrument.
Experiment template files and data files can be transferred to the desktop software.
Email notifications are not available.
Separate profiles are required for each instrument if multiple instruments are used.
Thermo Fisher
Connect profile
Experiment template files and data files are stored on the instrument.
Experiment template files and data files can be transferred to the Thermo Fisher
Connect
platform or the desktop software.
Email notifications are available.
The same profile can be used for multiple instruments.
Chapter6Configure the instrument and manage instrument profiles
Link an instrument profile to Thermo Fisher
Connect
6
QuantStudio
3 and 5 RealTime PCR Systems Installation, Use, and Maintenance Guide
87
Thermo Fisher
Connect Platform instrument profile roles and functions
The first user who links their local instrument profile to their Thermo Fisher
Connect Platform account
is assigned a Connect Platform profile with the administrator role.
Instrument
profile
Location Functions allowed
Standard Connect
Platform
Create, save, open, import, and run template files
Create and modify run settings
View and export data files
Administrator Connect
Platform
All the permissions of a local administrator profile, plus the following functions
performed in the Connect Platform:
See a list of all the Connect Platform profiles that are linked to the
instrument
Assign Connect Platform administrator roles to one or more users
Remove a user from an instrument
Disconnect the instrument from the Connect Platform
Change the instrument name
Link the instrument to your Thermo Fisher
Connect Platform account
This section describes using your Thermo Fisher
Connect Platform account when you use the
instrument for the first time.
The first time that you use your Connect Platform account on an instrument, you will be prompted to
create a four-digit numerical PIN. This PIN is to use when signing in to the instrument with your Connect
Platform account. It will apply to all other instruments when you use your Connect Platform account.
This does not change the password when signing in to your Connect Platform account on a browser.
After the PIN is set up on the instrument, it must be changed with your Connect Platform account on a
browser, see “Create an administrator instrument profile during initial start-up” on page85.
The first time the instrument is used with a Connect Platform account, a region must be selected.
1.
In the Sign In screen, tap Get started4Create a local instrument file4Connect to your
Thermo Fisher
Connect Platform account.
2.
(Optional) Select the appropriate region.
Option
Description
China For users in China
U.S. For users in any country other than China
Chapter6Configure the instrument and manage instrument profiles
Link an instrument profile to Thermo Fisher
Connect
6
88
QuantStudio
3 and 5 RealTime PCR Systems Installation, Use, and Maintenance Guide
3.
Tap a connection option.
Option Action
Mobile
devices
Note: Before selecting this option, install and sign in to the Connect Platform
application on your mobile device.
On the instrument:
a.
Tap
Mobile devices.
b.
Hold the camera on your mobile device over the QR code that is displayed on the
touchscreen.
c.
Tap Close.
PC A link code is displayed on the instrument.
On a computer:
a.
Access the Connect Platform.
b.
Click Add instrument.
c.
Select QuantStudio.
d.
Enter the link code.
4.
(Optional) In the Enter PIN screen, tap the PIN (4 digits required) field, enter a four-digit numerical
PIN, then tap Enter.
Tap the Show PIN checkbox to show or hide the PIN.
5.
(Optional) Tap the Confirm PIN field, enter the four-digit numerical PIN again, then tap Enter.
6.
Tap Done.
Link a local profile to a Thermo Fisher
Connect profile
This section describes linking a local instrument profile to a Thermo Fisher
Connect profile, if a local
instrument profile was created first.
The first time that you use your Thermo Fisher
Connect account on an instrument, you are prompted
to create a four-digit numerical PIN. This PIN is to use when signing in to the instrument with your
Thermo Fisher
Connect account. It applies to all other instruments when you use your Thermo Fisher
Connect account. This does not change the password when signing in to your Thermo Fisher
Connect
account on a browser.
After the PIN is set up on the instrument, it must be changed with your Thermo Fisher
Connect
account on a browser, see “Change the PIN for a Thermo Fisher
Connect profile” on page92.
The first time the instrument is used with a Thermo Fisher
Connect account, a region must be
selected.
Chapter6
Configure the instrument and manage instrument profiles
Link an instrument profile to Thermo Fisher
Connect
6
QuantStudio
3 and 5 RealTime PCR Systems Installation, Use, and Maintenance Guide
89
For a detailed description of the profile, see “If you link when you are signed in to the instrument” on
page91.
1.
In the home screen, tap (Profile).
The My Profile screen is displayed.
2.
Tap Connect4Link Account.
The Connect to Connect Platform screen is displayed.
3.
(Optional) Select the appropriate region.
Option Description
China For users in China
U.S. For users in any country other than China
4.
Tap a connection option.
Option
Action
Mobile
devices
Note: Before selecting this option, install and sign in to the Thermo Fisher
Connect
application on your mobile device.
On the instrument:
a.
Tap
Mobile devices.
b.
Hold the camera on your mobile device over the QR code that is displayed on the
touchscreen.
c.
Tap Close.
PC A link code is displayed on the instrument.
On a computer:
a.
Access the Thermo Fisher
Connect platform.
b.
Click Add instrument.
c.
Select QuantStudio.
d.
Enter the link code.
5.
(Optional) In the Enter PIN screen, tap the PIN (4 digits required) field, enter a four-digit numerical
PIN, then tap Enter.
Tap the Show PIN checkbox to show or hide the PIN.
6.
(Optional) Tap the Confirm PIN field, enter the four-digit numerical PIN again, then tap Enter.
7.
Tap Done.
Chapter6
Configure the instrument and manage instrument profiles
Link an instrument profile to Thermo Fisher
Connect
6
90
QuantStudio
3 and 5 RealTime PCR Systems Installation, Use, and Maintenance Guide
Unlink a Thermo Fisher
Connect account
Unlinking a Thermo Fisher
Connect is done from the Thermo Fisher
Connect platform on your
computer.
For more information about unlinking a Thermo Fisher
Connect account, see “If you link when you are
signed in to the instrument” on page91.
1.
Sign in to your account on the desktop Thermo Fisher
Connect platform.
2.
In the left pane, click
(Instrument).
3.
Select the instrument, then click Disconnect.
4.
Tap Confirm.
If you link when you are signed in to the instrument
In this scenario, your local instrument profile name is created manually on the instrument before you
link. Your local instrument profile name diers from your Thermo Fisher
Connect instrument profile
name.
Phase Steps that occur
Before you
link:
You enter your local instrument profile name in the Sign In screen.
Your local instrument profile (UserABC) is displayed in the home screen of the instrument.
All plates and results that you create are accessible only when you are signed in with your local
instrument profile.
When you
link:
You link your local instrument profile (see “Link a local profile to a Thermo Fisher
Connect
profile” on page89).
If this is the first time you link, a Thermo Fisher
Connect instrument profile is created using the
FirstNameLastInitial of the user name from your thermofisher.com account.
Example: User1@thermofisher.com First name is User, Last name is Gray. Thermo Fisher
Connect account username is UserG.
Your local instrument profile (UserABC) is linked to your Thermo Fisher
Connect account
(
User1@thermofisher.com).
Your Thermo Fisher
Connect instrument profile (UserG.) replaces your local instrument
profile.
After you link:
Your Thermo Fisher
Connect instrument profile (UserG.) and is displayed in the home
screen of the instrument.
Plate files and data files from your local instrument profile can be transferred to Thermo Fisher
Connect.
New plate files and data files are saved under your Thermo Fisher
Connect instrument profile.
Chapter6Configure the instrument and manage instrument profiles
Link an instrument profile to Thermo Fisher
Connect
6
QuantStudio
3 and 5 RealTime PCR Systems Installation, Use, and Maintenance Guide
91
(continued)
Phase Steps that occur
If your
Thermo
Fisher
Connect
account is
unlinked:
Your local instrument profile (UserABC) is displayed in the home screen of the instrument.
Plate files and data files that were saved under your Thermo Fisher
Connect instrument profile
are accessible under your local instrument profile.
Plate files and data files are saved under local instrument profile and can be copied to Thermo
Fisher
Connect (see “Manage templates (EDT files)” on page47 and “Transfer EDS files from
the instrument home screen” on page46).
Your local instrument profile name (UserABC) is available for selection in the Sign In screen.
Change the PIN for a Thermo Fisher
Connect profile
This section describes changing a PIN for a Thermo Fisher
Connect profile. To change a PIN for a
local instrument profile, see “Manage all instrument profiles (Administrator only)” on page85.
1.
Sign in to Thermo Fisher
Connect on a browser.
2.
Click Update PIN number.
3.
In the Update PIN number dialog box, enter a new PIN, then enter the new PIN a second time to
confirm it.
4.
Click Send.
The 4-digit PIN to sign in to instruments with your Thermo Fisher
Connect profile is updated. This
change applies to all of the instruments that you access with your Thermo Fisher
Connect profile.
Enable SAE mode (Administrator only)
Security, Audit, and e-Signature (SAE) functions are available only for the QuantStudio
5 RealTime
PCR System.
The SAE module works with the instrument and desktop software to record and restrict various user
activities. Audits and e-Signatures are tracked in the desktop software.
In the home screen:
1.
Touch Settings4Instrument Settings4SAE Mode.
2.
Slide the control On to enable SAE on the instrument.
3.
Touch OK.
Chapter6
Configure the instrument and manage instrument profiles
Enable SAE mode (Administrator only)
6
92
QuantStudio
3 and 5 RealTime PCR Systems Installation, Use, and Maintenance Guide
The following instrument functions are disabled in SAE mode.
Start a run from the instrument.
Note: All instrument runs must be started from desktop software that has SAE functions enabled.
Edit an instrument name.
Edit an instrument date or time.
Restore to factory defaults.
Perform a software update.
Access the instrument settings if you are not signed in.
Require instrument profile sign-in (Administrator only)
1.
In the home screen, touch Settings4Manage Users4Sign In Required.
2.
Slide the control On.
Only signed-in users are allowed to access the instrument for any task, including access to
Settings.
3.
Touch Done.
Enable remote instrument monitoring (Administrator only)
1.
In the home screen, touch Settings4Maintenance and Service4Monitoring.
2.
Configure the instrument monitoring settings.
Option
Description
Remote
Monitoring
Service
Slide the control On to enable the Remote Monitoring Service.
Enabling the Remote Monitoring Service allows the instrument to automatically
notify Thermo Fisher Scientific support teams in real time of potential instrument
issues. The service monitors and sends general instrument data, but the service
does not monitor or send
customer data.
Thermo Fisher
Connect Monitor
Slide the control On to enable users to access realtime monitoring of
amplification plots from the Thermo Fisher
Connect dashboard (cloud-based
platform).
3.
Touch OK.
Chapter6Configure the instrument and manage instrument profiles
Require instrument profile sign-in (Administrator only)
6
QuantStudio
3 and 5 RealTime PCR Systems Installation, Use, and Maintenance Guide
93
Update instrument software (Administrator only)
1.
In the home screen, touch Settings4Maintenance and Service4Software Update.
2.
Select the location of the update files.
3.
When prompted, confirm your request to update the software.
Restore factory defaults
This procedure can only be performed by an Administrator.
IMPORTANT!
Back up the instrument before restoring factory defaults (see “Back up or restore the
instrument” on page74).
1.
In the home screen, touch Settings4Instrument Settings4Restore Factory Defaults.
2.
Touch Restore Factory Defaults.
3.
Power O, then power On the instrument to apply the change.
Instrument profiles and files stored on the instrument are deleted, including all EDS files and
any custom dye calibrations.
The most recent valid ROI/Uniformity, Background, and system dye calibrations are not
deleted.
After restoring factory defaults, perform initial startup tasks. See “Initial startup” on page84.
Configure the network
IMPORTANT!
Configure the instrument for either wired connection or wireless connection but not
both.
“Set up a wired connection” on page95
“Set up a wireless connection” on page95
Chapter6Configure the instrument and manage instrument profiles
Update instrument software (Administrator only)
6
94
QuantStudio
3 and 5 RealTime PCR Systems Installation, Use, and Maintenance Guide
Set up a wired connection
1.
In the home screen, touch Settings4Instrument Settings4Network Connection4 Wired.
2.
In the Network Configuration screen, touch a radio button to connect to a network either by
DHCP or by a Static IP address.
Option Action
Automatic Touch DHCP.
Manual
a.
Touch Static IP.
b.
Enter the IP address for the instrument, the Subnet Mask, the Default Gateway, and
(optional) the Primary DNS Server and the Secondary DNS Server.
Addresses are in the form of X.X.X.X, where each X is a number from 0 to255.
Note: Ask your system administrator if the IP address is assigned statically or dynamically. For
static addresses, you need to know the IP address for the instrument, the Subnet Mask, and the
Default Gateway.
3.
Touch OK.
Set up a wireless connection
1.
In the home screen, touch Settings4Instrument Settings4Network Connection4
Wireless to display a list of the available networks.
2.
Select a network or touch Join other network, enter the network password, then touch Enter.
If you select Join other network, enter the network name and security type.
3.
Touch Join to continue.
4.
In the Network Connection Complete screen, touch Next.
5.
Touch OK.
6.
In the Network Configuration screen, edit the Network details or touch Done.
Chapter6Configure the instrument and manage instrument profiles
Configure the network
6
QuantStudio
3 and 5 RealTime PCR Systems Installation, Use, and Maintenance Guide
95
Select a region for the Thermo Fisher
Connect Platform
(Administrator only)
During the initial startup, you are automatically prompted to select a region for the Thermo Fisher
Connect Platform. Return to the selection through Settings, if necessary.
1.
In the home screen, touch Settings4Instrument Settings4Thermo Fisher Connect Region.
2.
Touch a region.
The server used to store your data is located in this region.
3.
Touch OK.
Manage the instrument name (Administrator only)
1.
In the home screen, touch Settings4Instrument Settings4Instrument Name.
2.
Touch the Instrument Name field, enter an instrument name, then touch Done.
3.
Touch OK.
Set the date and time (Administrator only)
The date and time cannot be set if the instrument is linked to the Thermo Fisher
Connect Platform.
1.
In the home screen, touch Settings4Instrument Settings4Date/Time.
2.
Select a time zone from the dropdown list.
3.
Select a date format.
a.
Touch Date Format, then select the preferred date format.
b.
Touch Next, touch the date field, then enter the date.
c.
Touch Enter, then touch Done.
4.
Select a time format.
a.
Touch Time Format.
b.
Slide the control to select a 12hour or 24hour clock.
c.
Touch Next, touch the time field, then enter the time.
d.
Touch Enter, then touch Done.
5.
Touch Done.
Chapter6
Configure the instrument and manage instrument profiles
Select a region for the Thermo Fisher
Connect Platform (Administrator only)
6
96
QuantStudio
3 and 5 RealTime PCR Systems Installation, Use, and Maintenance Guide
Manage the signout timer (Administrator only)
1.
In the home screen, touch Settings4Manage Users4Sign Out Timer.
2.
Touch the Edit Time field, then enter the desired duration of inactivity before automatic user sign
out.
3.
Touch Enter, then touch Done.
Chapter6Configure the instrument and manage instrument profiles
Manage the signout timer (Administrator only)
6
QuantStudio
3 and 5 RealTime PCR Systems Installation, Use, and Maintenance Guide
97
Install and connect the instrument to
a network
Workflow: Install and connect to anetwork .............................................. 98
Before youbegin ..................................................................... 99
Unpack and install theinstrument ...................................................... 99
Power on and follow the startup wizard ................................................ 100
Connect the computer to the instrument directly or to aLAN .............................. 100
Instrument and computerconnections ................................................. 101
Download and install the desktop software ............................................. 102
Networking ......................................................................... 103
Workflow: Install and connect to a network
Perform all steps in
“Before you begin” on page99
“Unpack and install the instrument” on page99
“Power on and follow the startup wizard” on page100
“Perform instrument verification using RNase P plates” on page59
“Connect the computer to the instrument directly or to a LAN” on page100
A
98
QuantStudio
3 and 5 RealTime PCR Systems Installation, Use, and Maintenance Guide
Before you begin
This section provides instructions for customer installation of the computer and the instrument. For
installation by a field service engineer (FSE), contact Support to order a service call.
Before starting the installation, complete the following tasks:
Determine the configuration for your instrument and obtain the network information you need (see
“Networking” on page103).
Review site requirements (see QuantStudio
3 and 5 RealTime PCR Systems Site Preparation
Guide (Pub. No.MAN0010405).
Review instrument and computer connections (see “Instrument and computer connections” on
page101).
Unpack and install the instrument
1.
Prepare the installation site as described in the QuantStudio
3 and 5 RealTime PCR Systems Site
Preparation Guide (Pub. No.MAN0010405).
2.
Follow the pre-printed instructions on the instrument box to unpack the instrument, accessories,
and reference documentation. Save the packing material for future use or recycle it.
The instrument box contains:
Reference documentation: Welcome note, unpacking and set up instructions card, system
documentation insert
One instrument
Accessories: power cable, Ethernet cable, USB drive, reaction tube retainer
Shipping plate
Note: Save the shipping plate but do not use it to operate the instrument.
If you ordered the wireless adapter, it is provided separately.
3.
Place the instrument on the bench.
4.
Plug the power cable into the power port on the back panel of the instrument, then plug the cable
into an electrical receptacle.
5.
Connect your instrument as required by your network configuration (see “Supported options for
instrument and computer connections” on page103 and “Instrument and computer connections”
on page101).
Connect an Ethernet cable to the Ethernet port on the back panel of the instrument, then
connect the cable to a computer or a networking port.
or
Connect wirelessly via the wireless adapter.
AppendixA
Install and connect the instrument to a network
Before you begin
A
QuantStudio
3 and 5 RealTime PCR Systems Installation, Use, and Maintenance Guide
99
Note: Do not connect the High Power USB WiFi Module (Cat. No. A26774) to the instrument if
it is connected to a network by an Ethernet cable. Configuring the instrument for both wired and
wireless connection can interfere with instrument operation.
Power on and follow the startup wizard
1.
Power on the instrument.
2.
Follow the startup wizard through the following tasks:
Select the language for the instrument.
Accept the license agreement.
Select a Wired or Wireless networking option, as needed for your configuration (see
“Supported options for instrument and computer connections” on page103).
Configure the instrument date and time.
Create an administrator instrument profile.
Note: You can perform any of the steps above at a later time if you do not have the information
needed to complete the startup screens. See “Configure instrument settings” on page27.
IMPORTANT!
Before using the instrument for the first time, we recommend that you perform
instrument verification using RNase P plates (see “Perform instrument verification using RNaseP
plates” on page59).
Connect the computer to the instrument directly or to a
LAN
This section describes the direct wired connection of the computer to the instrument or to a LAN.
1.
Connect an Ethernet cable from the instrument or a LAN to the computer.
2.
Power on the computer, then log in using a Windows
Administrator account.
3.
Open the Control Panel, then open the Network and Sharing Center.
4.
In the left pane of the Network and Sharing Center window, click Change Adapter Settings.
5.
Right-click on Broadband Connection, then select Properties.
6.
In the Networking tab, select Internet Protocol Version 4 (TCP/IPv4).
7.
Click Properties.
AppendixA
Install and connect the instrument to a network
Power on and follow the startup wizard
A
100
QuantStudio
3 and 5 RealTime PCR Systems Installation, Use, and Maintenance Guide
8.
In the Internet Protocol Version 4 (TCP/IPv4) Properties dialog box, select one of the following
options.
Obtain an IP address automatically
Use the following IP address
9.
If Use the following IP address was selected, enter the IP address.
10.
Close all dialog boxes by clicking OK, then re-start the computer.
Instrument and computer connections
1
7
66
5
4
2
3
Figure3Instrument back panel
1
USB ports
2
WiFi USB port—Connect USB wireless adapter for
wireless network access (ordered separately)
3
Ethernet Port—RJ45 port for 100/1,000 Mbps Ethernet
communication with the instrument
4
RS232 Port—For service use only
5
Fuse Cover
6
Power Switch
7
Power Port—100 to 240 VAC
AppendixAInstall and connect the instrument to a network
Instrument and computer connections
A
QuantStudio
3 and 5 RealTime PCR Systems Installation, Use, and Maintenance Guide
101
3
1 1
2
4
Figure4Instrumenttocomputer connections (barcode scanner connected to the computer)  Minitower
configuration
1
Detachable power supply cord compatible with local power supply receptacle.
2
Connection between the computer and the instrument.
3
Connection between the computer and the monitor, keyboard, and mouse.
4
Connection between the computer and the (optional) handheld barcode scanner.
Download and install the desktop software
Computer requirements for the desktop software
If you purchased a computer provided by Thermo Fisher Scientific, you can install the desktop software
and use it to control the instrument.
This instrument uses QuantStudio
Design and Analysis Software v1.x. QuantStudio
Design and
Analysis Software v2.x can also be used to set up and analyze files.
Thermo Fisher Scientific does not provide technical support for customer-provided computers. Thermo
Fisher Scientific provides limited support for Thermo Fisher Scientificsupplied computers that have
been altered from their default configuration.
You can install the desktop software on a customer-provided computer and use the software to create
templates and analyze data. Minimum requirements for a customer-provided computer are:
Operating system—Windows
10 (64bit)
Processor—Intel
Core
or compatible
Memory—4 GB RAM
Hard drive—20 GB free space
Monitor—1280 × 1024 resolution
AppendixA
Install and connect the instrument to a network
Download and install the desktop software
A
102
QuantStudio
3 and 5 RealTime PCR Systems Installation, Use, and Maintenance Guide
Download the desktop software
1.
Sign in to your account at thermofisher.com.
Note: If you do not have an account, create an account.
2.
Go to thermofisher.com/quantstudio3-5softwaredownloads.
3.
Download the software and the example files.
Install the desktop software
1.
Use a Windows
Administrator account to log in to the computer on which you are installing the
desktop software.
2.
Unzip the downloaded software and example files.
3.
Double-click setup.exe.
4.
Follow the InstallShield Wizard prompts to install the software.
5.
Accept the License Agreement.
6.
Select Typical as the setup preference, then click Next.
7.
If you are installing the software on a computer provided by Thermo Fisher Scientific, install
the software on the D:\ drive. If you are installing on a customer-provided computer, install the
software in your preferred location.
8.
Click Finish.
Networking
IMPORTANT!
This section provides general networking information. It does not provide adequate
detail to integrate the instrument into all possible network architectures. Because a network may
contain advanced features (such as a firewall or network domains), we recommend that you consult
a network administrator before connecting the instrument to your laboratory network.
Supported options for instrument and computer connections
We support the following direct, networked (LAN– local area network), or Thermo Fisher
Connect
Platform configurations. Configurations other than those listed are not recommended. Select
a configuration that meets the needs of your laboratory's instrument, software, and workflow
requirements.
IMPORTANT!
Do not connect both an Ethernet cable and the WiFi module (Cat.No.A26774) to
the instrument. Configuring the instrument for both wired and wireless connection can interfere with
instrument operation.
AppendixAInstall and connect the instrument to a network
Networking
A
QuantStudio
3 and 5 RealTime PCR Systems Installation, Use, and Maintenance Guide
103
Note: For detailed information about networking your instrument, see the QuantStudio
3 and 5
RealTime PCR Systems IT Checklist (Pub.No.MAN0013738).
Direct configuration option
A computer provided by Thermo Fisher Scientific with the QuantStudio
Design and Analysis Software
Computertoinstrument connection:
Direct, wired connection between the computer and the instrument using an
Ethernet cable
IMPORTANT! A direct instrumenttocomputer connection cannot be combined with the
LAN or the Thermo Fisher
Connect Platform configurations.
Networked configuration option
A computer provided by Thermo Fisher Scientific with the QuantStudio
Design and Analysis Software
The computer and instrument must have the same subnet mask within the same network.
ComputertoLAN connection (select an option):
Wired connection to the network using an Ethernet cable – or
Wireless connection to the network
LAN
LAN
InstrumenttoLAN connection (select an option):
Wired connection to the network using an Ethernet cable – or
Wireless connection to the network using the instrument WiFi module
Connect Platform configuration option
Internet access and a computer with the Chrome
web browser to access the Thermo Fisher
Connect
Platform
ComputertoThermo Fisher
Connect Platform connection (select an option):
Wired connection to the internet using an Ethernet cable – or
Wireless connection to the internet
LAN
InstrumenttoThermo Fisher
Connect Platform connection (select an option):
Wired connection to the network using an Ethernet cable – or
Wireless connection to the network using the instrument WiFi module
AppendixAInstall and connect the instrument to a network
Networking
A
104
QuantStudio
3 and 5 RealTime PCR Systems Installation, Use, and Maintenance Guide
Control and monitor networked instruments
When the instrument is connected to a network:
Computers on the network that are running the desktop software can control the instrument.
Networked instruments can be controlled by only one computer at a time.
Instruments linked to Thermo Fisher
Connect Platform cannot be controlled remotely. However,
you can perform the following functions:
Remotely access Thermo Fisher
Connect Platform to create (and analyze) experiments.
From the instrument, download the experiments and start a run.
Monitor a run from Thermo Fisher
Connect Platform in real time.
Ethernet port overview
The Ethernet port of the instrument supports:
Static IP network service with subnet mask, primary and secondary data network service (DNS),
and default gateway settings, or dynamic host configuration protocol (DHCP) network service.
mDNS/DNS for local domains.
Note: Because mDNS is limited to direct network connections, an instrument configured for
mDNS may not be visible to other nodes that are separated by a router, hub, or another network
device.
IPv4 linklocal (IPV4LL) in the RFC (also known as Automatic Private IP Addressing [APIPA] or
Internet Protocol Automatic Configuration [IPAC]).
Note: When an instrument is set for DHCP, APIPA is automatically enabled, and the instrument
provides an IP address when no address is supplied by the DHCP server.
Firewall ports that must be open
Ports
Condition
80/443
Standard ports for instrument-to-Thermo Fisher
Connect Platform and computer-to-Thermo
Fisher
Connect Platform connections
mDNS, 7443 Instrument-to-computer connection
mDNS, 5353 Instrument discovery
AppendixAInstall and connect the instrument to a network
Networking
A
QuantStudio
3 and 5 RealTime PCR Systems Installation, Use, and Maintenance Guide
105
Networking guidelines and best practices
Consult a network administrator before connecting the instrument to a network.
To enable the full functionality of the software, the computer requires a network connection.
Open the firewall port for the instruments to be discovered. See “Firewall ports that must be open”
on page105.
Observe the restrictions to mDNS and Autodiscovery.
The instrument supports mDNS but only when the instrument and computer share a direct network
connection and are within the same subnet. Network computers that are separated from the
instrument by a router, hub, or another network device may not be able to access the instrument by
its host name.
Confirm the uniqueness of the instrument name.
The instrument name must be unique within the subnet. The desktop software can
automatically discover instruments on the link-local network.
The instrument does not test the uniqueness of the instrument name within the subnet when it
is set.
AppendixAInstall and connect the instrument to a network
Networking
A
106
QuantStudio
3 and 5 RealTime PCR Systems Installation, Use, and Maintenance Guide
Troubleshooting
Observation Possible cause Recommended action
Inconsistent communication
between instrument and
computer or instrument and
Thermo Fisher
Connect
Platform
The instrument is configured
for both wired and wireless
network connection.
Ensure only one connectivity option is plugged
into the instrument (either an Ethernet cable or a
wireless adapter, but not both).
Configure for wired or wireless network connection.
Weak or unstable internet
connection, especially if
configured for wireless.
Change the configuration to a wired connection.
Use a wireless network with a stronger or more
consistent signal.
The connection between the
instrument and the computer
is not recognized
The connection is not fully
established.
Power the instrument o, then power it on again.
If using a networked
configuration, the instrument
and computer are not on the
same subnet mask.
Contact your information technologies department
to have them ensure that the instrument and
computer are on the same subnet mask.
If using a networked
configuration, the instrument
or computer has an invalid
IP address.
Contact your information technologies department
to have them ensure that the IP addresses are
valid.
Insucient disk space
message
Insucient disk space to
save a run.
1.
In the home screen, touch Settings4Run
History4Manage.
2.
Delete or transfer experiments from the
instrument.
The touchscreen is black
The instrument is in sleep
mode.
Touch anywhere on the instrument touchscreen.
The instrument is not
powered on.
If you touch the instrument touchscreen and it
remains black, check if the instrument is powered
on. The power switch is located on the rear panel
of the instrument.
If the instrument does not power on, check that the
instrument is properly plugged in.
If the instrument does not power on and the
instrument is properly plugged in, contact Support.
Forgot PIN for instrument
profile
Non-administrator forgot
instrument profile PIN.
See “Manage all instrument profiles (Administrator
only)” on page85.
B
QuantStudio
3 and 5 RealTime PCR Systems Installation, Use, and Maintenance Guide
107
Observation Possible cause Recommended action
Forgot PIN for instrument
profile
(continued)
Administrator forgot
instrument profile PIN.
Have another administrator reset the PIN for the
forgotten-PIN profile (see “Manage all instrument
profiles (Administrator only)” on page85).
If there is not another administrator profile on the
instrument, you must restore factory defaults (see
“Restore factory defaults” on page94).
AppendixBTroubleshooting
Networking
B
108
QuantStudio
3 and 5 RealTime PCR Systems Installation, Use, and Maintenance Guide
Parts and materials
Kits, consumables, accessories, and reagents .......................................... 109
Consumables (96well, 0.2mLformat) ................................................. 109
384-wellconsumables ............................................................... 110
Consumables (96-well, 0.1-mLformat) ................................................. 111
Accessories ........................................................................ 112
General-use materials andconsumables ............................................... 112
Kits, consumables, accessories, and reagents
Unless otherwise indicated, all materials are available through thermofisher.com.
Store all calibration and RNase P plates at –20°C. All other items can be stored at 15–30°C. Use all
materials by the expiration date on the packaging.
Catalog numbers that appear as links open the web pages for those products.
Consumables (96well, 0.2mL format)
Consumable
Amount Cat.No.
MicroAmp
Optical 8-Cap Strips 300 strips 4323032
MicroAmp
Optical 8-Tube Strip, 0.2mL 125 strips 4316567
MicroAmp
Optical Tube without Cap, 0.2mL 2,000 tubes N8010933
MicroAmp
96-Well Tray/Retainer Set for Veriti
Systems 10 trays 4381850
MicroAmp
Optical 96-Well Reaction Plate
10 plates N8010560
500 plates 4316813
MicroAmp
Optical 96-Well Reaction Plate with Barcode
20 plates 4306737
500 plates 4326659
MicroAmp
EnduraPlate
Optical 96-Well Clear Reaction Plates with Barcode
20 plates 4483354
500 plates 4483352
MicroAmp
EnduraPlate
Optical 96-Well Clear GPLE Reaction Plates with
Barcode
20 plates 4483348
C
QuantStudio
3 and 5 RealTime PCR Systems Installation, Use, and Maintenance Guide
109
(continued)
Consumable
Amount Cat.No.
MicroAmp
EnduraPlate
Optical 96-Well Clear GPLE Reaction Plates with
Barcode
500 plates 4483351
MicroAmp
Optical Adhesive Film Kit 1 kit 4313663
MicroAmp
Optical Film Compression Pad (Optional, used for some applications) 5 each 4312639
Instrument verification or calibration plate Cat.No.
TaqMan
RNase P Instrument Verification Plate, 96Well 0.2mL 4432382
Region of Interest (ROI) and Background Plates, 96Well 0.2mL (2 plates) 4432364
QuantStudio
3/5 10Dye Spectral Calibration Kit, 96Well 0.2mL
(contains all 3 spectral calibration plates listed below)
A26343
QuantStudio
3/5 Spectral Calibration Plate 1, 96Well 0.2mL
(FAM
,VIC
,ROX
,SYBR
dyes)
A26331
QuantStudio
3/5 Spectral Calibration Plate 2, 96Well 0.2mL
(ABY
,JUN
,MUSTANG PURPLE
dyes)
A26332
QuantStudio
3/5 Spectral Calibration Plate 3, 96Well 0.2mL
(TAMRA
,NED
,Cy
®
5dyes)
A26333
384-well consumables
Consumable
Contents Cat.No.
MicroAmp
Optical 384-Well Reaction Plate with Barcode 50 plates 4309849
500 plates 4326270
1000 plates 4343814
MicroAmp
EnduraPlate
Optical 384-Well Reaction Plate with Barcode (clear) 20 plates 4483285
500 plates 4483273
MicroAmp
Optical Adhesive Film Kit 1 kit 4313663
AppendixCParts and materials
384-well consumables
C
110
QuantStudio
3 and 5 RealTime PCR Systems Installation, Use, and Maintenance Guide
Calibration or instrument verification plate Cat.No.
384-Well Region of Interest (ROI) and Background Plates (2 plates) 4432320
QuantStudio
3 or 5 10-Dye Spectral Calibration Kit, 384-Well A26341
384-Well Spectral Calibration Plate 1
(containing FAM
, VIC
, ROX
, TAMRA
, and SYBR
dyes)
A26334
384-Well Spectral Calibration Plate 2
(containing ABY
, JUN
, MUSTANG PURPLE
, NED
, and Cy
®
5 dyes)
A26335
384-Well TaqMan
RNase P Instrument Verification Plate 4455280
Consumables (96-well, 0.1-mL format)
Consumable
Contents Cat.No.
MicroAmp
Optical 8-Cap Strip 300 strips 4323032
MicroAmp
Optical Fast 8-Tube Strip (0.1 mL) 125 strips 4358293
MicroAmp
Optical Fast Tube with Cap (0.1 mL) 1000 tubes 4358297
MicroAmp
96-Well Tray for VeriFlex
Block 10 trays 4379983
MicroAmp
Optical 96-Well Fast Reaction Plate (0.1 mL) 10 plates 4346907
MicroAmp
EnduraPlate
Optical 96-Well Fast Reaction Plate with Barcode (clear)
(0.1 mL)
20 plates 4481194
500 plates 4483494
MicroAmp
EnduraPlate
Optical 96-Well Fast Clear Reaction Plates with Barcode 20 plates 4483485
MicroAmp
Optical Adhesive Film Kit 1 kit 4313663
Calibration or instrument verification plate Cat.No.
96-Well Fast Region of Interest (ROI) and Background Plates (2 plates) 4432426
QuantStudio
3 or 5 10-Dye Spectral Calibration Kit, 96-Well 0.1-mL A26342
96-Well 0.1-mL Spectral Calibration Plate 1
(containing FAM
, VIC
, ROX
, and SYBR
dyes
A26336
96-Well 0.1-mL Spectral Calibration Plate 2
(containing ABY
, JUN
, and MUSTANG PURPLE
dyes)
A26337
96-Well 0.1-mL Spectral Calibration Plate 3
(containing TAMRA
, NED
, and Cy
®
5 dyes)
A26340
96Well Fast TaqMan
RNase P Instrument Verification Plate 4351979
AppendixCParts and materials
Consumables (96-well, 0.1-mL format)
C
QuantStudio
3 and 5 RealTime PCR Systems Installation, Use, and Maintenance Guide
111
Accessories
Item
Amount Cat.No.
MicroAmp
Multi Removal Tool 1 tool 4313950
MicroAmp
Cap Installing Tool (handle style) 1 tool 4330015
MicroAmp
Optical Adhesive Film 25 films 4360954
100 films 4311971
MicroAmp
Adhesive Film Applicator 5 applicators 4333183
RT-PCR Grade Water 10×1.5 mL tubes AM9935
Handheld Barcode Scanner 1 scanner 4488442
High Power USB WiFi Module 1 module A26774
General-use materials and consumables
The following general-use materials and consumables are required to calibrate, maintain, and operate
the instrument.
Unless otherwise indicated, all materials are available through thermofisher.com. "MLS" indicates that
the material is available from fisherscientific.com or another major laboratory supplier.
Material/Consumable
Source
Centrifuge with 96well plate buckets MLS
Cotton or nylon swabs and lintfree cloths MLS
Ethanol, 95% solution MLS
Optical clear adhesive film for PCR MLS
Na-hypochlorite, 0.1% v/v solution MLS
Pipettors, 100µL and 200µL (with pipette tips) MLS
Powderfree gloves MLS
Safety glasses MLS
Screwdriver, flathead MLS
Tissue, lintfree MLS
Deionized water MLS
AppendixCParts and materials
Accessories
C
112
QuantStudio
3 and 5 RealTime PCR Systems Installation, Use, and Maintenance Guide
Instrument specification and layout
Configured systemdimensions ........................................................ 114
Electrical requirements ............................................................... 115
Environmental requirements .......................................................... 116
Network requirements ............................................................... 116
D
QuantStudio
3 and 5 RealTime PCR Systems Installation, Use, and Maintenance Guide
113
Configured system dimensions
Allow space for the configured instrument. A typical setup with a co-located minitower computer is
shown below.
Note: Dimensions are rounded to the nearest whole or half unit.
15 cm (6 in.)
27 cm (10.5 in.)
50 cm (20 in.)
30 cm (12 in.)
40 cm (16 in.)
207 cm (81.5 in.)
AppendixDInstrument specification and layout
Configured system dimensions
D
114
QuantStudio
3 and 5 RealTime PCR Systems Installation, Use, and Maintenance Guide
Instrument clearances
During instrument installation and maintenance, it is necessary to access the back of the instrument. If
the back of the instrument faces a wall, ensure that there is sucient clearance on the bench to rotate
the instrument for access.
IMPORTANT! For safety, the power outlet for the instrument must be accessible.
Component Top Front Sides Back
Instrument 30 cm (12in.) 30 cm (12in.) 15 cm (6in.) 15 cm (6in.)
Computer
[1]
15 cm (6in.) 15 cm (6in.)
[1]
Co-locating the computer with the instrument is optional.
Electrical requirements
WARNING!
For safety, the power outlet used for powering the instrument must be accessible at all
times. See “Instrument clearances” on page115 for information about the space needed between the
wall and the instrument. In case of emergency, you must be able to immediately disconnect the main
power supply to all the equipment. Allow adequate space between the wall and the equipment so
that the power cords can be disconnected in case of emergency.
WARNING! Use of accessories, transducers, and cables other than those specified or provided by
the manufacturer of the equipment could result in increased electromagnetic emissions or decreased
electromagnetic immunity of this equipment and result in improper operation.
Electric receptacle with grounding capability
Maximum power dissipation: ~960 W (not including computer and monitor)
Mains AC line voltage tolerances must be up to ±10 percent of nominal voltage
Device
Rated voltage Circuit required Rated frequency Rated power
Instrument 100–240 ±10% VAC
[1]
10 A 50/60 Hz 960 W
Computer (laptop) 100–240 ±10% VAC 10 A 50/60 Hz 90 VA
Computer (desktop)
100–240 ±10% VAC 10 A 50/60 Hz
125 VA
Monitor 65 VA
[1]
If the supplied power fluctuates beyond the rated voltage, a power line regulator may be required. High or low voltages can adversely
affect the electronic components of the instrument.
AppendixD
Instrument specification and layout
Electrical requirements
D
QuantStudio
3 and 5 RealTime PCR Systems Installation, Use, and Maintenance Guide
115
Environmental requirements
Table2Environmental requirements
Condition Acceptable range
Installation site Indoor use only
Electromagnetic
interference
Do not use this device in close proximity to sources of strong electromagnetic radiation
(for example, unshielded intentional RF sources). Strong electromagnetic radiation may
interfere with the proper operation of the device.
Altitude Between sea level and 2000m (6500ft) above sea level
Operating conditions
Humidity: 15–80% relative humidity (noncondensing)
Temperature: 15°C to 30°C (59°F to 86°F)
Note: For optimal performance, avoid rapid or extreme fluctuations in room
temperature.
Storage and transport
conditions
Humidity: 20–80% relative humidity (noncondensing)
Temperature: –30°C to 60°C (–22°F to 140°F)
Thermal output During operation, the net thermal output, based on the actual current draw of the
instrument, is expected to be approximately 960 W (3275Btu/h).
Vibration Ensure that the instrument is not adjacent to strong vibration sources, such as a
centrifuge, pump, or compressor. Excessive vibration will aect instrument performance.
Pollution degree The instrument has a Pollution Degree rating of II. The instrument may only be installed
in an environment that has nonconductive pollutants such as dust particles or wood
chips. Typical environments with a Pollution Degree II rating are laboratories and sales and
commercial areas.
The noise output of the instrument is ≤60 dB when running.
Other conditions Ensure the instrument is located away from any vents that could expel particulate material
onto the instrument components.
Avoid placing the instrument and computer adjacent to heaters, cooling ducts, or in direct
sunlight.
Network requirements
The instrument:
Is factory-configured for IPv4 TCP/IP communication and includes an Ethernet adapter
(100/1,000 Mbps) with an RJ45-type connector for integrating the device into a local area network
(LAN).
Can alternatively be configured for wireless networking (High Power USB WiFi Module required,
sold separately as an optional accessory).
The instrument can be configured for either wired or wireless networking, not both.
AppendixD
Instrument specification and layout
Environmental requirements
D
116
QuantStudio
3 and 5 RealTime PCR Systems Installation, Use, and Maintenance Guide
If a Thermo Fisher Scientific service representative is to install the instrument:
If the instrument will be connected to a LAN, an active, tested network jack must be in place before
the scheduled installation date.
A representative from your information technologies department must be available during the
installation to help connect the instrument to your network.
Required materials to network the instrument:
Wired—Ethernet cable of sucient length with RJ45 connectors
CAT5 cable for a 100Mbps network connection
CAT5e or CAT6 cable for a 1,000Mbps network connection
or
Wireless—High Power USB WiFi Module (Cat.No.A26774, sold separately)
AppendixDInstrument specification and layout
Network requirements
D
QuantStudio
3 and 5 RealTime PCR Systems Installation, Use, and Maintenance Guide
117
Safety
Symbols on thisinstrument ........................................................... 118
Safety alerts on thisinstrument ....................................................... 121
Safety information for instruments not manufactured by Thermo FisherScientific ............ 122
Instrumentsafety .................................................................... 122
Safety and electromagnetic compatibility (EMC) standards ............................... 124
Chemicalsafety ..................................................................... 127
Biological hazardsafety .............................................................. 128
WARNING!
GENERAL SAFETY. Using this product in a manner not specified in the user
documentation may result in personal injury or damage to the instrument or device. Ensure that
anyone using this product has received instructions in general safety practices for laboratories and
the safety information provided in this document.
·
Before using an instrument or device, read and understand the safety information provided in the
user documentation provided by the manufacturer of the instrument or device.
·
Before handling chemicals, read and understand all applicable Safety Data Sheets (SDSs) and use
appropriate personal protective equipment (gloves, gowns, eye protection, and so on). To obtain
SDSs, visit thermofisher.com/support.
Symbols on this instrument
Symbols may be found on the instrument to warn against potential hazards or convey important safety
information. In this document, the hazard symbol is used along with one of the following user attention
words:
CAUTION!—Indicates a potentially hazardous situation that, if not avoided, may result in minor or
moderate injury. It may also be used to alert against unsafe practices.
WARNING!—Indicates a potentially hazardous situation that, if not avoided, could result in death or
serious injury.
DANGER!—Indicates an imminently hazardous situation that, if not avoided, will result in death or
serious injury.
Symbol
English Français
Caution, risk of danger
Consult the manual for further safety
information.
Attention, risque de danger
Consulter le manuel pour d’autres
renseignements de sécurité.
E
118
QuantStudio
3 and 5 RealTime PCR Systems Installation, Use, and Maintenance Guide
(continued)
Symbol English Français
Caution, risk of electrical shock Attention, risque de choc électrique
Moving parts Parties mobiles
Caution, hot surface Attention, surface chaude
Potential biohazard Danger biologique potentiel
Ultraviolet light Rayonnement ultraviolet
Potential slipping hazard Danger de glisser potentiel
On On (marche)
O O (arrêt)
On/O On/O (marche/arrêt)
Standby En attente
Earth (ground) terminal Borne de (mise à la) terre
Protective conductor terminal (main
ground)
Borne de conducteur de protection (mise à la
terre principale)
Terminal that can receive or supply
alternating current or voltage
Borne pouvant recevoir ou envoyer une
tension ou un courant de type alternatif
Terminal that can receive or supply
alternating or direct current or voltage
Borne pouvant recevoir ou envoyer une
tension ou un courant continu ou alternatif
AppendixESafety
Symbols on this instrument
E
QuantStudio
3 and 5 RealTime PCR Systems Installation, Use, and Maintenance Guide
119
(continued)
Symbol English Français
Do not dispose of this product in
unsorted municipal waste
CAUTION! To minimize
negative environmental impact
from disposal of electronic
waste, do not dispose of
electronic waste in unsorted
municipal waste. Follow local
municipal waste ordinances for
proper disposal provision and
contact customer service for
information about responsible
disposal options.
Ne pas éliminer ce produit avec les déchets
usuels non soumis au tri sélectif.
MISE EN GARDE! Pour minimiser
les conséquences négatives sur l’en
vironnement à la suite de l’élimination
de déchets électroniques, ne pas éli
miner ce déchet électronique avec les
déchets usuels non soumis au tri sé
lectif. Se conformer aux ordonnances
locales sur les déchets municipaux
pour les dispositions d’élimination et
communiquer avec le service à la cli
entèle pour des renseignements sur
les options d’élimination responsable.
Conformity symbols
Conformity mark
Description
Indicates conformity with safety requirements for Canada and U.S.A.
Indicates conformity with European Union requirements for safety and electromagnetic
compatibility.
Indicates conformity with Australian standards for electromagnetic compatibility.
AppendixESafety
Symbols on this instrument
E
120
QuantStudio
3 and 5 RealTime PCR Systems Installation, Use, and Maintenance Guide
Safety alerts on this instrument
Additional text may be used with one of the symbols described above when more specific information
is needed to avoid exposure to a hazard. See the following table for safety alerts found on the
instrument.
English Français
CAUTION! Hazardous chemicals.
Read the Safety Data Sheets (SDSs)
before handling.
MISE EN GARDE! Produits chimiques dange
reux. Lire les fiches signalétiques (FS) avant de ma
nipuler les produits.
CAUTION! Hazardous waste. Refer
to SDS(s) and local regulations for
handling and disposal.
MISE EN GARDE! Déchets dangereux. Lire les
fiches signalétiques (FS) et la réglementation locale
associées à la manipulation et à l’élimination des
déchets.
Location of safety labels on the instrument
AppendixESafety
Safety alerts on this instrument
E
QuantStudio
3 and 5 RealTime PCR Systems Installation, Use, and Maintenance Guide
121
Safety information for instruments not manufactured by
Thermo Fisher Scientific
Some of the accessories provided as part of the instrument system are not designed or built by Thermo
Fisher Scientific. Consult the manufacturer's documentation for the information needed for the safe use
of these products.
Instrument safety
General
CAUTION! Do not remove instrument protective covers. If you remove the protective instrument
panels or disable interlock devices, you may be exposed to serious hazards including, but not limited
to, severe electrical shock, laser exposure, crushing, or chemical exposure.
CAUTION! Solvents and Pressurized fluids. Wear eye protection when working with any
pressurized fluids. Use caution when working with any polymeric tubing that is under pressure:
·
Extinguish any nearby flames if you use flammable solvents.
·
Do not use polymeric tubing that has been severely stressed or kinked.
·
Do not use polymeric tubing with tetrahydrofuran or nitric and sulfuric acids.
·
Be aware that methylene chloride and dimethyl sulfoxide cause polymeric tubing to swell and
greatly reduce the rupture pressure of the tubing.
·
Be aware that high solvent flow rates (~40mL/min) may cause a static charge to build up on the
surface of the tubing and electrical sparks may result.
AppendixESafety
Safety information for instruments not manufactured by Thermo Fisher Scientific
E
122
QuantStudio
3 and 5 RealTime PCR Systems Installation, Use, and Maintenance Guide
Physical injury
CAUTION! Moving and Lifting Injury. The instrument is to be moved and positioned only by the
personnel or vendor specified in the applicable site preparation guide. Improper lifting can cause
painful and permanent back injury.
Things to consider before lifting or moving the instrument or accessories:
·
Depending on the weight, moving or lifting may require two or more persons.
·
If you decide to lift or move the instrument after it has been installed, do not attempt to do so
without the assistance of others, the use of appropriate moving equipment, and proper lifting
techniques.
·
Ensure you have a secure, comfortable grip on the instrument or accessory.
·
Make sure that the path from where the object is to where it is being moved is clear of
obstructions.
·
Do not lift an object and twist your torso at the same time. Keep your spine in a good neutral
position while lifting with your legs.
·
Participants should coordinate lift and move intentions with each other before lifting and carrying.
·
For smaller packages, rather than lifting the object from the packing box, carefully tilt the box on its
side and hold it stationary while someone else slides the contents out of the box.
CAUTION!
Moving Parts. Moving parts can crush, pinch and cut. Keep hands clear of moving parts
while operating the instrument. Disconnect power before servicing.
WARNING! Do not attempt to lift or move the instrument without the assistance of others. Use
appropriate moving equipment and proper lifting technique, improper lifting may result in serious
injury.
AppendixESafety
Instrument safety
E
QuantStudio
3 and 5 RealTime PCR Systems Installation, Use, and Maintenance Guide
123
Electrical safety
WARNING! Fuse Installation. Before installing the instrument, verify that the fuses are properly
installed and the fuse voltage matches the supply voltage. Replace fuses only with the type and
rating specified for the unit. Improper fuses can damage the instrument wiring system and cause a
fire.
WARNING! Ensure appropriate electrical supply. For safe operation of the instrument:
·
Plug the system into a properly grounded receptacle with adequate current capacity.
·
Ensure the electrical supply is of suitable voltage.
·
Never operate the instrument with the ground disconnected. Grounding continuity is required for
safe operation of the instrument.
WARNING! Power Supply Line Cords. Use properly configured and approved line cords for the
power supply in your facility.
WARNING! Disconnecting Power. To fully disconnect power either detach or unplug the power
cord, positioning the instrument such that the power cord is accessible.
Cleaning and decontamination
CAUTION! Cleaning and Decontamination. Use only the cleaning and decontamination methods
specified in the manufacturer's user documentation. It is the responsibility of the operator (or other
responsible person) to ensure the following requirements are met:
·
No decontamination or cleaning agents are used that could cause a HAZARD as a result of a
reaction with parts of the equipment or with material contained in the equipment.
·
The instrument is properly decontaminated a) if hazardous material is spilled onto or into the
equipment, and/or b) prior to having the instrument serviced at your facility or sending the
instrument for repair, maintenance, trade-in, disposal, or termination of a loan (decontamination
forms may be requested from customer service).
·
Before using any cleaning or decontamination methods (except those recommended by the
manufacturer), users should confirm with the manufacturer that the proposed method will not
damage the equipment.
Safety and electromagnetic compatibility (EMC) standards
The instrument design and manufacture complies with the following standards and requirements for
safety and electromagnetic compatibility.
AppendixE
Safety
Safety and electromagnetic compatibility (EMC) standards
E
124
QuantStudio
3 and 5 RealTime PCR Systems Installation, Use, and Maintenance Guide
Safety compliance
Reference
Description
EU Directive 2014/35/EU European Union “Low Voltage Directive”
IEC 61010-1
EN 61010-1
UL 61010-1
CSA C22.2 No. 61010-1
Safety requirements for electrical equipment for measurement, control, and
laboratory use – Part 1: General requirements
IEC 61010-2-010
EN 61010-2-010
Safety requirements for electrical equipment for measurement, control and
laboratory use – Part 2-010: Particular requirements for laboratory equipment for
the heating of materials
IEC 61010-2-081
EN 61010-2-081
Safety requirements for electrical equipment for measurement, control and
laboratory use – Part 2-081: Particular requirements for automatic and semi-
automatic laboratory equipment for analysis and other purposes
EMC
Reference
Description
Directive 2014/30/EU European Union “EMC Directive”
EN 61326-1/ IEC 61326-1 Electrical Equipment for Measurement, Control and Laboratory Use –
EMC Requirements – Part 1: General Requirements
AS/NZS CISPR 11 Limits and Methods of Measurement of Electromagnetic Disturbance
Characteristics of Industrial, Scientific, and Medical (ISM) Radiofrequency
Equipment
ICES-001, Issue 4 Industrial, Scientific and Medical (ISM) Radio Frequency Generators
FCC Part 15 Subpart B (47 CFR) U.S. Standard Radio Frequency Devices
This equipment has been tested and found to comply with the limits for a Class A digital device,
pursuant to part 15 of the FCC Rules. These limits are designed to provide reasonable protection
against harmful interference when the equipment is operated in a commercial environment. This
equipment generates, uses, and can radiate radio frequency energy and, if not installed and used
in accordance with the instruction manual, may cause harmful interference to radio communications.
Operation of this equipment in a residential area is likely to cause harmful interference in which case the
user will be required to correct the interference at his own expense.
AppendixESafety
Safety and electromagnetic compatibility (EMC) standards
E
QuantStudio
3 and 5 RealTime PCR Systems Installation, Use, and Maintenance Guide
125
Environmental design
Reference
Description
Directive 2012/19/EU European Union “WEEE Directive” – Waste electrical and
electronic equipment
EU Directive 2011/65/EU & Commission
Delegated Directive (EU) 2015/863
European Union “RoHS Directive” – Restriction of hazardous
substances in electrical and electronic equipment
Directive 2006/66/EC European Union “Battery Directive”
China EEP Hazardous Substances Information
Component Name QuantStudio
3 and 5 RealTime PCR Systems
Hazardous Substances
(Pb) (Hg) (Cd) (Cr(VI)) (PBB) (PBDE)
PCBA X O O O O O
SJ/T11364. This table is compiled according to SJ/T 11364 standard.
O: GB/T26572.
Indicates that the concentration of the hazardous substance in all homogeneous materials for the part is
below the relevant threshold of the GB/T 26572 standard.
X: GB/T26572.
Indicates that the concentration of the hazardous substance in at least one homogenous material of the
part is above the relevant threshold of the GB/T 26572 standard.
AppendixESafety
Safety and electromagnetic compatibility (EMC) standards
E
126
QuantStudio
3 and 5 RealTime PCR Systems Installation, Use, and Maintenance Guide
Chemical safety
WARNING! GENERAL CHEMICAL HANDLING. To minimize hazards, ensure laboratory personnel
read and practice the general safety guidelines for chemical usage, storage, and waste provided
below. Consult the relevant SDS for specific precautions and instructions:
·
Read and understand the Safety Data Sheets (SDSs) provided by the chemical manufacturer
before you store, handle, or work with any chemicals or hazardous materials. To obtain SDSs, see
the "Documentation and Support" section in this document.
·
Minimize contact with chemicals. Wear appropriate personal protective equipment when handling
chemicals (for example, safety glasses, gloves, or protective clothing).
·
Minimize the inhalation of chemicals. Do not leave chemical containers open. Use only with
sucient ventilation (for example, fume hood).
·
Check regularly for chemical leaks or spills. If a leak or spill occurs, follow the manufacturer
cleanup procedures as recommended in the SDS.
·
Handle chemical wastes in a fume hood.
·
Ensure use of primary and secondary waste containers. (A primary waste container holds the
immediate waste. A secondary container contains spills or leaks from the primary container.
Both containers must be compatible with the waste material and meet federal, state, and local
requirements for container storage.)
·
After emptying a waste container, seal it with the cap provided.
·
Characterize (by analysis if needed) the waste generated by the particular applications, reagents,
and substrates used in your laboratory.
·
Ensure that the waste is stored, transferred, transported, and disposed of according to all local,
state/provincial, and/or national regulations.
·
IMPORTANT! Radioactive or biohazardous materials may require special handling, and disposal
limitations may apply.
WARNING!
HAZARDOUS WASTE (from instruments). Waste produced by the instrument is
potentially hazardous. Follow the guidelines noted in the preceding General Chemical Handling
warning.
WARNING! 4L Reagent and Waste Bottle Safety. Four-liter reagent and waste bottles can crack
and leak. Each 4-liter bottle should be secured in a low-density polyethylene safety container with the
cover fastened and the handles locked in the upright position.
AppendixESafety
Chemical safety
E
QuantStudio
3 and 5 RealTime PCR Systems Installation, Use, and Maintenance Guide
127
Biological hazard safety
WARNING! Potential Biohazard. Depending on the samples used on this instrument, the surface
may be considered a biohazard. Use appropriate decontamination methods when working with
biohazards.
WARNING! BIOHAZARD. Biological samples such as tissues, body fluids, infectious agents,
and blood of humans and other animals have the potential to transmit infectious diseases.
Conduct all work in properly equipped facilities with the appropriate safety equipment (for example,
physical containment devices). Safety equipment can also include items for personal protection,
such as gloves, coats, gowns, shoe covers, boots, respirators, face shields, safety glasses, or
goggles. Individuals should be trained according to applicable regulatory and company/ institution
requirements before working with potentially biohazardous materials. Follow all applicable local,
state/provincial, and/or national regulations. The following references provide general guidelines when
handling biological samples in laboratory environment.
·
U.S. Department of Health and Human Services, Biosafety in Microbiological and Biomedical
Laboratories (BMBL), 6th Edition, HHS Publication No. (CDC) 300859, Revised June 2020
www.cdc.gov/labs/pdf/CDC-BiosafetymicrobiologicalBiomedicalLaboratories-2020-P.pdf
·
Laboratory biosafety manual, fourth edition. Geneva: World Health Organization; 2020 (Laboratory
biosafety manual, fourth edition and associated monographs)
www.who.int/publications/i/item/9789240011311
AppendixESafety
Biological hazard safety
E
128
QuantStudio
3 and 5 RealTime PCR Systems Installation, Use, and Maintenance Guide
Documentation and support
Related documentation
Document
Publication
number
Description
QuantStudio
Design and
Analysis Desktop Software
Command-Line Application Guide
MAN0010409 Describes how to use the command-line software
application and provides the procedure to automate
the creation of new experiment files and export data
from existing files.
SAE Administrator Console v2.1
User Guide
MAN0028414 Describes how to use the SAE Administrator
Console.
QuantStudio
3 and 5 RealTime
PCR Systems Site Preparation
Guide
MAN0010405 Explains how to prepare your site to receive and
install the system. Intended for personnel who
schedule, manage, and perform the tasks required
to prepare the site for installation.
Note: For additional documentation, see “Customer and technical support” on page129.
Obtain information from the Help system
The instrument has a Help system that describes how to use each feature of the touchscreen. Touch
on the instrument touchscreen to access the Help system.
Customer and technical support
Visit thermofisher.com/support for the latest service and support information.
Worldwide contact telephone numbers
Product support information
Product FAQs
Software, patches, and updates
Training for many applications and instruments
Order and web support
F
QuantStudio
3 and 5 RealTime PCR Systems Installation, Use, and Maintenance Guide
129
Product documentation
User guides, manuals, and protocols
Certificates of Analysis
Safety Data Sheets (SDSs; also known as MSDSs)
Note: For SDSs for reagents and chemicals from other manufacturers, contact the
manufacturer.
Limited product warranty
Life Technologies Corporation and/or its aliate(s) warrant their products as set forth in the
Life Technologies' General Terms and Conditions of Sale at www.thermofisher.com/us/en/home/
global/terms-and-conditions.html. If you have any questions, please contact Life Technologies at
www.thermofisher.com/support.
AppendixFDocumentation and support
Limited product warranty
F
130
QuantStudio
3 and 5 RealTime PCR Systems Installation, Use, and Maintenance Guide
Index
A
accessories 112
administrator instrument profile 85
Administrator profile 88
B
background calibration
create plate 58
description 50
plate, prepare 53
Background calibration
run 54
when to perform 49
backup instrument 12, 74
barcode scan 36
biohazard safety 128
C
calibration
reminders 51
Background 54
custom dye. See custom dye
custom melt 72
Dye 54
plate, prepare 53
reminders 51
ROI/Uniformity 54
schedule 49
status 51
transfer results 56
view images 56
workflow 52
calibration failure, troubleshoot
57
Change password 92
Change PIN 92
Cloud server, select location 96
comparative Ct runs 19
computer
connection to instrument or LAN 103
connections 101
requirements for desktop software 102
set up 100
Connect
Link to instrument 88
Roles
Administrator 88
Standard 88
Unlink from instrument 91
Connect profile, Overview 87
connect to network, workflow 98
connection, not recognized
consumables
384-well 110
96-well 0.1-mL 111
96-well 0.2-mL 109
contamination, identify 57
custom dye
add to the instrument 69
calibration plate, create 68
calibration, perform 70
description 14
dilution guidelines 65
optimal concentration 68
wavelength requirements 65
custom dye calibration, workflow 65
custom dye dilution plate, prepare 66
custom melt calibration 72
D
data collection 15
data collection points, add to method 37
date and time 96
delete files 47
desktop software
download 103
download and install 102
install 103
disk space message
documentation, related 129
download desktop software 103
dye calibration
description 50
plate, prepare 53
Dye calibration
run 54
when to perform 49
QuantStudio
3 and 5 RealTime PCR Systems Installation, Use, and Maintenance Guide
131
dyes
custom 14, 65
system 13
E
Edit profile 92
EDS files, transfer 46
EDT files, copy or delete 47
Ethernet port 105
experiment
create from template 33
edit 35
run from saved file 33
run last 34
F
factory defaults 94
firewall ports 105
fluorescence high
fuses, replace 80
G
genotyping runs 19
guest
profile
disable 93
limits 26
H
Help system, access 129
home screen, parts of 24
I
Index Term 43, 75
install, desktop software 103
installation
before you begin 99
desktop software 102
workflow 98
instrument
profile, create 26
connections 101
enable for monitoring 93
move 82
name 96
overview 12
parts of 16
power o 82
power on 23, 81
prepare to store or move or ship 82
return for service 83
touchscreen black
instrument filters 13
instrument layout 113
instrument settings 27
instrument specification 113
instrument verification. See RNase P verification
L
limited product warranty 130
load plate 40, 41
Local instrument profile, Overview 87
lock the touchscreen 46
log records, view 75
M
maintenance 74
melt curve runs 19
melt curves, add to method 37
method
add melt curves 37
add stages 37
add steps 37
data collection points 37
edit 36
parts of 37
pause 39
ramp rate 39
VeriFlex Zone 39
N
network
configurations supported 103
guidelines and best practices 106
wired connection set up 95
wireless connection set up 95
network configuration and security 11
networking 103
O
online Help. See Help system
Order to set up profiles 87
P
Password, Change 92
password security 11
pause run 45
performance verification. See RNase P verification
PIN, Change 92
Index
132
QuantStudio
3 and 5 RealTime PCR Systems Installation, Use, and Maintenance Guide
PIN reset 85
presence/absence runs 19
Profile
Connect
Link 88
Unlink 91
Edit 92
profiles
administrator 85
delete 85
edit 86
guest, disable 93
instrument 26
require sign-in 93
reset PIN 85
sign in 27
Profiles
Connect
Administrator 88
Standard 88
Order to set up 87
properties, enter or edit 35
protocol. See method
R
real-time plots
view on instrument touchscreen 45
zoom 46
related documentation 129
relative standard curve runs 19
Replace with index term
require sign in 93
restore instrument 12, 74
RNase P
verification
pass criteria 61
plate description 59
run 62
RNaseP verification, plate, prepare 61
ROI/uniformity calibration
description 50
plate, prepare 53
ROI/Uniformity calibration
run 54
when to perform 49
run
pause or stop 45
post-PCR 15
real-time 15
run data, transfer 46
run history 47
run last experiment 34
run protocol. See method
runs, types 19
S
SAE mode (Security, Audit, and eSignature) 92
safety, biohazard 128
sample block
clean 78
decontaminate 77
settings 27
sign in, require 93
sign out
profile 27
timer, set 97
sleep mode 81
software features 16
software updates 94
stages, add to method 37
standard curve runs 19
Standard profile 88
standby mode 81
startup wizard 100
steps, add to method 37
stop run 45
support, customer and technical 129
supported dyes 13
symbols, safety 118
system dyes 13
T
template properties. See properties
terms and conditions 130
thermal protocol. See method
Thermo Fisher Connect profile, Edit 92
third-party software 19
transfer files 47
troubleshooting 107
U
uniformity calibration
description 50
plate, prepare 53
unload plate 40, 41
unpack instrument 99
use the instrument, workflow 32
V
verification failure, troubleshoot 63
VeriFlex Zone, add to method 39
VeriFlex Zones, block description 15
Index
QuantStudio
3 and 5 RealTime PCR Systems Installation, Use, and Maintenance Guide
133
W
warranty 130
well details
define, assign, view 39
view on instrument touchscreen 45
workflow, use the instrument 32
Z
zoom 46
Index
134
QuantStudio
3 and 5 RealTime PCR Systems Installation, Use, and Maintenance Guide
QS 35 Instrument UG_RUO_MAN0010407-v16-GUID-88D256DF-7F65-4A04-
B662-0DCAA6DC61A6-2024/07/02 20:58:23 en
16:45:28.621+01:00
thermofisher.com/support|thermofisher.com/askaquestion
thermofisher.com
8 August 2024